Device evaluated by mfr.: synergy ous mr 2.75 x 16mm stent delivery system catheter was returned for analysis.Visual, tactile and microscopic analysis were performed on the device.Examination of the hypotube found a multiple kinks and breaks at 14cm, 25.7cm and 40.9cm distal to the distal end of the strain relief, 100.5cm from the distal markerband.Multiple kinks were identified along the polymer extrusion shaft.Examination of the stent found no issues.There was no sign of damage, stretching or lifting of the stent struts.Balloon cones were reviewed, and no issues were noted.Balloon wings were tightly wrapped and evenly folded and were not subjected to positive pressure.No signs of movement, stent was set between the proximal and distal markerbands.Bumper tip showed no signs of distal tip damage.No other device issues were identified during returned product analysis.
|