MAUDE Adverse Event Report: SYNTHES GMBH DRILL BIT Ø2.8 QC L135 CALIBRATION 45; BIT, DRILL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SYNTHES GMBH DRILL BIT Ø2.8 QC L135 CALIBRATION 45; BIT, DRILL Back to Search Results
Catalog Number 03.133.106S
Device Problem Device-Device Incompatibility (2919)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 11/27/2023
Event Type  malfunction  
Event Description
Device report from japan reports an event as follows: it was reported that on (b)(6) 2023, the patient underwent orif with the locking screw and the drill guide in question for a humeral distal fracture.Drilling was performed to insert the locking screw into the plate shaft.During drilling, the drill bit got stuck in the drill guide and could not be pulled out.The procedure was completed successfully with no adverse patient impact.There were no fragments generated, no additional medical intervention was necessary.No additional information is available.This report is for a 2.8mm drill bit qc 200mm 110mm calibration.This is report 2 of 2 for (b)(4).
 
Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.D9: complainant part is not expected to be returned for manufacturer review/investigation.E3: initial reporter is a synthes employee.H3, h4, h6: without a lot number the device history records review could not be completed.Product was not returned.Based on the information available, it has been determined that no corrective and/or preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.H10 additional narrative: e1: initial reporter's phone number added.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.D1: device brand name updated.D2a: common device name updated.D2b: procode updated.D4: catalog number updated.D4: udi number updated.
 
Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.H3, h6: the device was received, the investigation is in progress, no conclusion could be drawn at the time of filing this report.If information is obtained that was not available for this medwatch, a follow-up medwatch will be filed as appropriate.
 
Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.Visual analysis of the returned sample revealed that the drill bit ø2.8 qc l135 calibration 45.Was unable to disassemble with the threaded drill guide 2.8 f/screw 3.5 f/v.A dimensional inspection was not performed for the device due to post manufacturing damage.The observed condition of the device was consistent with a random component failure that may have been caused by exposure to unintended forces.As part of depuy synthes quality process, all devices are manufactured, inspected, and released to approved specifications.The overall complaint was confirmed as the observed condition of the device would contribute to the complained device issue.There is no indication that a design or manufacturing issue has caused the complaint condition.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.If information is obtained that was not available for this medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DRILL BIT Ø2.8 QC L135 CALIBRATION 45
Type of Device
BIT, DRILL
Manufacturer (Section D)
SYNTHES GMBH
eimattstrasse 3
oberdorf 4436
SZ  4436
Manufacturer (Section G)
MONUMENT
1101 synthes avenue
monument CO 80132
Manufacturer Contact
kate karberg
1302 wright lane east
west chester, PA 19380
3035526892
MDR Report Key18347038
MDR Text Key331073193
Report Number8030965-2023-15846
Device Sequence Number1
Product Code HTW
UDI-Device Identifier07612334150042
UDI-Public(01)07612334150042
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number03.133.106S
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 11/27/2023
Initial Date FDA Received12/18/2023
Supplement Dates Manufacturer Received12/21/2023
02/02/2024
02/16/2024
Supplement Dates FDA Received01/11/2024
02/14/2024
02/20/2024
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? Yes
Type of Device Usage Unknown
Patient Sequence Number1
Treatment
THREADED DRILL GUIDE 2.8 F/SCREW 3.5 F/V; UNK - SCREWS: LOCKING
-
-