• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZIMMER BIOMET, INC. PT HYBRID GLEN POST REGENEREX; SHOULDER, PROSTHESIS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER BIOMET, INC. PT HYBRID GLEN POST REGENEREX; SHOULDER, PROSTHESIS Back to Search Results
Catalog Number PT-113950
Device Problem Fracture (1260)
Patient Problem Muscle/Tendon Damage (4532)
Event Date 11/22/2023
Event Type  Injury  
Manufacturer Narrative
(b)(4).Multiple mdr reports were filed for this event, please see associated reports: 0001825034 -2023 -03008.D10: sm hybrid glenoid base 4mm cat: 113952 lot: 346830.G2: foreign: australia.Customer has indicated that the product is in process of being returned to zimmer biomet for investigation.Once the investigation has been completed, a follow-up report will be submitted.H3 other text : product not returned.
 
Event Description
It was reported patient was revised due to rotator cuff failure and implant fracture approximately eight months post operation.The glenoid post and poly were exchanged.No additional patient consequences were reported.Attempts have been made and additional information on the reported event is unavailable at this time.
 
Manufacturer Narrative
This follow-up report is being submitted to relay additional information.D10: versa-dial 42x18x46 hum head; cat: 113032; lot: 293080.Compr nano hmrl pps 30mm; cat: 115730; lot: 271400.Versa-dial/comp ti std taper; cat: 118001; lot: 386740.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event Description
It was reported patient underwent a revision procedure approximately 21 months post implantation due to rotator cuff failure and glenoid post implant fracture.The humeral and glenoid components were removed and exchanged without complication.No additional patient consequences were reported.
 
Manufacturer Narrative
This follow-up report is being submitted to relay additional information.Visual examination of the returned product/provided pictures identified signs of use with some debris around the threads of the glenoid base as well as some scratches, gouges, and debris embedded into the working surface of the glenoid base.The threaded posts also have damage near the top of the threads.The regenerex post is fractured and not all of the post was returned.Fracture analysis could not be performed as the fracture is too smeared.Device history record was reviewed and no discrepancies were found.Radiographs were provided and reviewed by a health care professional.Review of the available records identified the following: two views of the right shoulder demonstrate a shoulder arthroplasty with radiolucent glenoid component and what appears to be a fractured central peg.Humeral component is intact with radiolucency along the baseplate.Superior subluxation of the humeral head is noted on the second image.The reported event is confirmed.Root cause was unable to be determined.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event Description
No further event information available at the time of this report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PT HYBRID GLEN POST REGENEREX
Type of Device
SHOULDER, PROSTHESIS
Manufacturer (Section D)
ZIMMER BIOMET, INC.
56 e. bell drive
p.o. box 587
warsaw IN 46581
Manufacturer (Section G)
ZIMMER BIOMET, INC.
56 e. bell drive
p.o. box 587
warsaw IN 46581
Manufacturer Contact
jennifer rapsavage
56 e. bell dr.
warsaw, IN 46582
5745260384
MDR Report Key18348093
MDR Text Key330799452
Report Number0001825034-2023-03007
Device Sequence Number1
Product Code HSD
UDI-Device Identifier00880304459854
UDI-Public(01)00880304459854(17)290102(10)187380
Combination Product (y/n)N
Reporter Country CodeAS
PMA/PMN Number
K060694
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 04/11/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberPT-113950
Device Lot Number187380
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 11/22/2023
Initial Date FDA Received12/18/2023
Supplement Dates Manufacturer Received04/01/2024
Supplement Dates FDA Received04/11/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured01/02/2019
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
SEE H10.; SEE H10.
Patient Outcome(s) Required Intervention; Hospitalization;
Patient SexFemale
Patient Weight65 KG
-
-