MEDTRONIC PUERTO RICO OPERATIONS CO. ACTIVA; STIMULATOR, ELECTRICAL, IMPLANTED, FOR PARKINS
|
Back to Search Results |
|
Model Number 37612 |
Device Problems
Failure to Power Up (1476); Charging Problem (2892); Connection Problem (2900)
|
Patient Problems
Muscular Rigidity (1968); No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 12/12/2023 |
Event Type
malfunction
|
Manufacturer Narrative
|
Continuation of d10: brand name restore; product id 37751 (serial: (b)(6); product type: 0213-recharger.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
It was reported that the patient wasn't able to charge their implant battery and the patient was "really stiff" because their battery hadn't been charged for 3 days and was about to die.Patient representative (rep) had said the reason the patient wasn't able to charge was the desktop charger cord was damaged and couldn't charge the recharger.Patient rep said the recharging port of the recharger was damaged and the recharger won't turn on.The recharger antenna cord was also damaged.Patient services reviewed wireless recharger replacement process, the patient rep requested that the legacy equipment be replaced with legacy equipment so that the patient could have replacement equipment.An email was sent to repair to replace the recharger, desktop charger, and recharger antenna.
|
|
Search Alerts/Recalls
|
|
|