• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ALIGN TECHNOLOGY, INC. INVISALIGN SYSTEM; ALIGNER, SEQUENTIAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ALIGN TECHNOLOGY, INC. INVISALIGN SYSTEM; ALIGNER, SEQUENTIAL Back to Search Results
Model Number INVISALIGN SYSTEM - COMPREHENSIVE
Device Problem Appropriate Term/Code Not Available (3191)
Event Date 11/15/2023
Event Type  Injury  
Event Description
The treating doctor reported that the patient had symptom of tooth extraction of tooth 2.7.It is unknown if the patient required any medical intervention to alleviate the reported symptom.It is unknown if the patient was prescribed any medication to alleviate the reported symptom.It is unknown if the patient is continuing the use of the aligners.
 
Manufacturer Narrative
The current instructions for use (ifu) contains the following: "precautions - a tooth that has been previously traumatized or significantly restored may be aggravated.In rare instances, the useful life of the tooth may be reduced, the tooth may require additional dental treatment such as endodontic and/or additional restorative work, and/or the tooth may be lost"; and "the health of the bone and gums which support the teeth may be impaired or aggravated"; and "the length of the roots of the teeth may be shortened during orthodontic treatment, which may become a threat to the longevity of the teeth".The potential root cause of this event is unknown.After reviewing available records (scan, photos, and x-rays) and per align's clinical assessment, the tooth #2.7 appeared to be considerably restored (almost the entire crown).Tooth #2.7 had a large resin with lack of anatomy, root canal treatment and possible tooth structure damage underneath.The mentioned clinical condition indicates a poor prognosis of the molar (#2.7).No conclusive evidence has been provided that supports or opposes the fact that the invisalign system aligners caused or contributed to the reported symptom of tooth loss.This event is being filed as an mdr as the patient reported tooth loss (serious injury) and the invisalign system aligners were being used.Note: due to the limited available information about this event, the date of the event is estimated to have occurred around 11/15/2023 (based on manufacturing date, delivery of aligners, and when the event was reported to align), as indicated in field b3; however, no conclusive evidence has been provided to confirm.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INVISALIGN SYSTEM
Type of Device
ALIGNER, SEQUENTIAL
Manufacturer (Section D)
ALIGN TECHNOLOGY, INC.
2820 orchard parkway
san jose CA 95134
Manufacturer (Section G)
ALIGN TECHNOLOGY, INC.
2820 orchard parkway
san jose CA 95134
Manufacturer Contact
harper shore
3030 slater road
morrisville, NC 27560
4084701343
MDR Report Key18350900
MDR Text Key330759099
Report Number2953749-2023-03573
Device Sequence Number1
Product Code NXC
UDI-Device Identifier00816063020189
UDI-Public(01)00816063020189(10)0166282893(13)231102(91)19143575XXN
Combination Product (y/n)N
Reporter Country CodeIT
PMA/PMN Number
K220287
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional
Reporter Occupation Physician
Type of Report Initial
Report Date 12/18/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Lay User/Patient
Device Model NumberINVISALIGN SYSTEM - COMPREHENSIVE
Device Catalogue Number9000
Device Lot Number166282893
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 11/23/2023
Initial Date FDA Received12/18/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured11/02/2023
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
Patient Outcome(s) Disability;
Patient Age31 YR
Patient SexMale
-
-