• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: OLYMPUS WINTER & IBE GMBH RESECTION SHEATH, 8 MM, FOR 8.5 MM/26 FR. OUTER SHEATH, ABS; HYSTEROSCOPE (AND ACCESSORIES)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

OLYMPUS WINTER & IBE GMBH RESECTION SHEATH, 8 MM, FOR 8.5 MM/26 FR. OUTER SHEATH, ABS; HYSTEROSCOPE (AND ACCESSORIES) Back to Search Results
Model Number A42011A
Device Problem Break (1069)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 11/27/2023
Event Type  malfunction  
Manufacturer Narrative
The subject device was returned to an olympus repair center for evaluation.The device evaluation found; the sealing ring was damaged, the tube was rusty and the ceramic tip was damaged.The investigation is ongoing.A supplemental report will be submitted upon completion of the investigation or if any additional information is provided.
 
Event Description
A user facility returned the olympus asset, resection sheath, 8 mm, for 8.5 mm/26 fr.Outer sheath, abs to the olympus service center.Upon inspection and testing of the returned device, it was observed that the ceramic tip was damaged.This report is being submitted for the reportable malfunction found during evaluation.There was no patient involvement.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RESECTION SHEATH, 8 MM, FOR 8.5 MM/26 FR. OUTER SHEATH, ABS
Type of Device
HYSTEROSCOPE (AND ACCESSORIES)
Manufacturer (Section D)
OLYMPUS WINTER & IBE GMBH
kuehnstrasse 61
hamburg, hamburg 22045
GM  22045
Manufacturer (Section G)
OLYMPUS WINTER & IBE GMBH
kuehnstrasse 61
hamburg, hamburg 22045
GM   22045
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
5082077661
MDR Report Key18352170
MDR Text Key330934969
Report Number9610773-2023-03682
Device Sequence Number1
Product Code HIH
UDI-Device Identifier04042761023658
UDI-Public04042761023658
Combination Product (y/n)N
Reporter Country CodeCH
PMA/PMN Number
K931995
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,User Facility
Reporter Occupation Other
Type of Report Initial
Report Date 12/18/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/18/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberA42011A
Device Lot Number20507
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/29/2023
Was the Report Sent to FDA? No
Date Manufacturer Received12/05/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured06/08/2020
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Treatment
ESG-400
-
-