Model Number 16-02-80
Device Problem Temperature Problem (3022)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 11/20/2023
Event Type  malfunction  
Manufacturer Narrative
A.1.-a.5.Patient information was not provided.G.5.The heater-cooler 16-02-80 is not distributed in the usa, but it is similar to heater-cooler 16-02-85, which is distributed in the usa (510(k) number: k191402).H10: livanova deutschland manufactures the heater-cooler system 3t.The incident occurred in turkey.If any additional information pertinent to the reported event is received, it will be provided in a supplemental report.
 
Event Description
Livanova deutschland received a report that, during a procedure, a heater-cooler system 3t displayed an error message (error 6) associated pump is using too much power.There is no report of any patient injury.The 3t heater cooler was inspected and it was found the distributor board and stirrer motor need to be replaced.
 
Event Description
See initial report.
 
Manufacturer Narrative
H10: complaints database analysis revealed no similar event since unit installation in 2022.Based on the information collected, it cannot be excluded that because of rusted/corroded/damaged motor bearings, likely due to environmental conditions, as water infiltration, humidity, condensation or high temperatures, the motor was no longer rotating freely and required a power overload to work, which could have led to an over-current that ultimately caused damages to the distribution board.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HEATER COOLER 3T (230 VOLT)
Type of Device
CONTROLLER, TEMPERATURE, CARDIOPULMONARY BYPASS
Manufacturer (Section D)
LIVANOVA DEUTSCHLAND GMBH
lindberghstrasse 25
munich
Manufacturer (Section G)
LIVANOVA DEUTSCHLAND
lindberghstrasse 25
munich 80939
GM   80939
Manufacturer Contact
enrico greco
14401 w. 65th way
arvada, CO 80004
MDR Report Key18355019
MDR Text Key330876649
Report Number9611109-2023-00643
Device Sequence Number1
Product Code DWC
UDI-Device Identifier04033817901099
UDI-Public(01)04033817901099(11)220221
Combination Product (y/n)N
Reporter Country CodeTU
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional
Reporter Occupation Other
Remedial Action Other
Type of Report Initial,Followup
Report Date 02/29/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number16-02-80
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 11/23/2023
Initial Date FDA Received12/19/2023
Supplement Dates Manufacturer Received01/31/2024
Supplement Dates FDA Received02/29/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured02/21/2022
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-