LIVANOVA DEUTSCHLAND GMBH HEATER COOLER 3T (230 VOLT); CONTROLLER, TEMPERATURE, CARDIOPULMONARY BYPASS
|
Back to Search Results |
|
Model Number 16-02-80 |
Device Problem
Temperature Problem (3022)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 11/20/2023 |
Event Type
malfunction
|
Manufacturer Narrative
|
A.1.-a.5.Patient information was not provided.G.5.The heater-cooler 16-02-80 is not distributed in the usa, but it is similar to heater-cooler 16-02-85, which is distributed in the usa (510(k) number: k191402).H10: livanova deutschland manufactures the heater-cooler system 3t.The incident occurred in turkey.If any additional information pertinent to the reported event is received, it will be provided in a supplemental report.
|
|
Event Description
|
Livanova deutschland received a report that, during a procedure, a heater-cooler system 3t displayed an error message (error 6) associated pump is using too much power.There is no report of any patient injury.The 3t heater cooler was inspected and it was found the distributor board and stirrer motor need to be replaced.
|
|
Event Description
|
See initial report.
|
|
Manufacturer Narrative
|
H10: complaints database analysis revealed no similar event since unit installation in 2022.Based on the information collected, it cannot be excluded that because of rusted/corroded/damaged motor bearings, likely due to environmental conditions, as water infiltration, humidity, condensation or high temperatures, the motor was no longer rotating freely and required a power overload to work, which could have led to an over-current that ultimately caused damages to the distribution board.
|
|