Model Number CI-1601-04 |
Device Problem
Adverse Event Without Identified Device or Use Problem (2993)
|
Patient Problems
Unspecified Infection (1930); Pain (1994)
|
Event Date 11/23/2023 |
Event Type
Injury
|
Manufacturer Narrative
|
Disclaimer: advanced bionics does not intend that this report be any admission of liability, fault or product defect.
|
|
Event Description
|
The recipient's device was reportedly explanted due to infection.The recipient was treated with antibiotics (type unknown).The recipient experience slight pain and irritation around site.The recipient will be reimplanted at a later date.The recipient is healing well after surgery.
|
|
Manufacturer Narrative
|
Advanced bionics considers the investigation into this reportable event as closed.The external visual inspection revealed the electrode was severed prior to receipt and silicone damage was observed on the top cover of the device.This is believed to have occurred during revision surgery, the device passed photographic imaging inspection.System lock was verified.The condition of the electrode prevented an electrical test from being performed.The device passed the electrical and mechanical tests performed.This device was explanted for medical reasons.The device passed the tests performed.This is the final report.Disclaimer: advanced bionics does not intend that this report be any admission of liability, fault or product defect.
|
|