This is a follow up report related to the same mdr event already included in the following report numbers filed with the fda: 3007797756-2023-00208, 3007797756-2023-00209, 3007797756-2023-00210 and 3007797756-2023-00211.This report covers the additional device used in this case.Pneumothorax is the most common side effect associated with the zephyr valve treatment (criner et al.A multicenter randomized controlled trial of zephyr endobronchial valve treatment in heterogeneous emphysema (liberate).Am j respir crit care med.2018; 198 (9): 1151-1164).Targeted lobar deflation likely causes inflation of the ipsilateral lobe, which can result in a tear of the already compromised parenchymal tissue of the emphysematous ipsilateral lobe, resulting in a pneumothorax (criner et al.A multicenter randomized controlled trial of zephyr endobronchial valve treatment in heterogeneous emphysema (liberate).Am j respir crit care med.2018; 198 (9): 1151-1164).In the liberate study (ide clinical study used to support pma p180002's approval), 26.6% of the zephyr valve subjects experienced a pneumothorax in the treatment period ([less than or equal to 45 days).These were managed using standard of care procedures as per previously published guidelines (valipour, arschang, et al.Respiration 87.6 (2014): 513-521).In 17.4% of the events, the pneumothorax resolved without any additional intervention with subjects under careful observation.In over half the events (56.5%), the pneumothorax was managed with a chest-tube only.An additional 13% of the events were managed with a chest-tube and the temporary removal of one or more valves, while another 13% of the events were managed with a chest-tube and removal of all the implanted valves.Upon successful resolution of the pneumothorax, removed valves can be replaced.Patients that experienced a pneumothorax experienced clinical benefits of the zephyr valve treatment that were similar to the benefits experienced by patients who did not have a pneumothorax.The zephyr ebv system ifu and pulmonx training program both specifically reference pneumothorax as a known side effect of this procedure and the published guidelines (valipour, arschang, et al."expert statement: pneumothorax associated with endoscopic valve therapy for emphysema-potential mechanisms, treatment algorithm, and case examples." respiration 87.6 (2014): 513-521).The reported event aligns with the experience observed in the liberate clinical study and is an expected side effect to the zephyr valve treatment.
|
See additional manufacture narrative section for more details.The patient had an endobronchial valve procedure on (b)(6) 2023.The patient presented with a pneumothorax on (b)(6) 2023, and a chest tube was placed on the same day to help with air-leak.The chest tube was removed on (b)(6) 2023.The patient presented with another pneumothorax on (b)(6) 2023, and a chest tube was placed on the same day to help with air-leak.The chest tube was removed on (b)(6) 2023.The patient presented with another pneumothorax on (b)(6) 2023, and a chest tube was placed on the same day to help with air-leak.All valves were removed on (b)(6) 2023.The chest tube was removed on (b)(6) 2023.The event was resolved on october 27, 2023, and the patient was discharged on (b)(6) 2023.
|