• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BECTON, DICKINSON AND CO. BD SYRINGE 10ML SALINE XS; SALINE VASCULAR ACCESS FLUSH

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BECTON, DICKINSON AND CO. BD SYRINGE 10ML SALINE XS; SALINE VASCULAR ACCESS FLUSH Back to Search Results
Catalog Number 306572
Device Problem Short Fill (1575)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 10/19/2023
Event Type  malfunction  
Event Description
It was reported that bd syringe 10ml saline xs did not have the designated amount of saline.The following information was provided by the initial reporter: the customer states that, "visited patient for disconnection.Syringe removed out of packaging for flushing purposes.On removal, noticed syringe had only 6ml instead of stated 10ml on packaging".
 
Manufacturer Narrative
H.3.A device evaluation and/or device history review is anticipated but is not complete.Upon completion a supplemental report will be filed.
 
Event Description
No additional information.
 
Manufacturer Narrative
A device history record review was completed for provided material number 306572 and lot number 3089846.The review did not reveal any possible non-conformances during the production process that could have contributed to this reported incident.To aid in the investigation of this issue, four (4) picture samples were received for evaluation by our quality team.The picture samples confirmed the defect of underfilled syringe.Although we were able to confirm the defect, an exact cause for the underfilled syringe could not be determined.If the physical sample becomes available, we would appreciate the opportunity to review it.This is the first report received for this type of issue on material (b)(4) and lot 3089846.At this time, further action has not been determined necessary.Our quality team will continue to closely monitor the manufacturing process for signs of this potential defect and any emerging trends.
 
Manufacturer Narrative
Correction made: we became aware on (b)(6) 2024 that the awareness date for this complaint was (b)(6) 2023.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD SYRINGE 10ML SALINE XS
Type of Device
SALINE VASCULAR ACCESS FLUSH
Manufacturer (Section D)
BECTON, DICKINSON AND CO.
donore road
drogheda
Manufacturer (Section G)
BECTON, DICKINSON AND CO.
donore road
drogheda
Manufacturer Contact
helen cox (mdr)
75 north fairway drive
vernon hills, IL 60061
8473935694
MDR Report Key18361405
MDR Text Key330983665
Report Number9616657-2023-00056
Device Sequence Number1
Product Code NGT
Combination Product (y/n)N
Reporter Country CodeUK
PMA/PMN Number
UKN
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup
Report Date 04/12/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number306572
Device Lot Number3089846
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 11/29/2023
Initial Date FDA Received12/19/2023
Supplement Dates Manufacturer Received02/26/2024
02/26/2024
Supplement Dates FDA Received02/26/2024
04/12/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/30/2023
Is the Device Single Use? Yes
Type of Device Usage Initial
Patient Sequence Number1
-
-