• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MAQUET CARDIOVASCULAR LLC VASOVIEW HEMOPRO 2; ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MAQUET CARDIOVASCULAR LLC VASOVIEW HEMOPRO 2; ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES Back to Search Results
Model Number VH-4000
Device Problems Smoking (1585); Environmental Particulates (2930); Material Twisted/Bent (2981)
Patient Problem Hemorrhage/Bleeding (1888)
Event Date 12/01/2023
Event Type  Injury  
Manufacturer Narrative
Tw id# (b)(4).The device has not yet been returned to maquet cardiac surgery for evaluation.We are following up with the customer for the return of the device.A supplemental report will be submitted if the device is received.
 
Event Description
The hospital reported that during an endoscopic vein harvesting procedure, vasoview hemopro 2 branch transection was complete.Harvester was then spot cauterizing side wall bleeding, a lot of smoke coming from the device.Upon removal of the device from the patient, the wire had lifted from the jaws.Power setting was at 2.5.A new device was used to complete the procedure.There was no delay.No patient harm was reported.
 
Manufacturer Narrative
Trackwise#: (b)(4).Corrected sections: h6--medical device ¿ problem code corrected from "1585" to "2930." the lot # 3000349829 history record review was completed.There were no ncmrs, rework, or deviations documented for the reported lot number.Based on the dhr/lhr review results, it was determined that there is no relation between the batch manufacturing process and the reported failure.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VASOVIEW HEMOPRO 2
Type of Device
ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES
Manufacturer (Section D)
MAQUET CARDIOVASCULAR LLC
45 barbour pond drive
wayne NJ
Manufacturer (Section G)
MAQUET CARDIOVASCULAR LLC
45 barbour pond drive
wayne NJ
Manufacturer Contact
arelean guzman
45 barbour pond drive
wayne, NJ 
MDR Report Key18364429
MDR Text Key330996153
Report Number2242352-2023-01045
Device Sequence Number1
Product Code GEI
UDI-Device Identifier00607567700406
UDI-Public00607567700406
Combination Product (y/n)N
PMA/PMN Number
K101274
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility,Company Representative
Reporter Occupation Physician Assistant
Type of Report Initial,Followup
Report Date 01/15/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/20/2023
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberVH-4000
Device Catalogue NumberVH-4000
Device Lot Number3000349829
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received01/15/2024
Was Device Evaluated by Manufacturer? No
Date Device Manufactured10/24/2023
Is the Device Single Use? Yes
Type of Device Usage Initial
Patient Sequence Number1
Treatment
UNKNOWN.
Patient Outcome(s) Required Intervention;
-
-