(b)(4).Date sent: 12/20/2023.D4: batch # a9df01.Additional information was requested and the following was obtained: "please provide more details ¿the clip could not be clipped deeply¿ did device was difficult to fire? yes, device was difficult to fire.Did device not fire clips (jammed)?device did not fire clips.Did device fire malformed clips? no.Did device fire scissored clips? no.If other, please specify.Device was difficult to fire and did not fire clips properly.The handle of the devise was hard.Detail information may be added" "the clip malformed in tear drop shape.The surgeon commented that the trigger was not grasped enough." investigation summary: the product was returned to ethicon for evaluation.Visual inspection and functional testing were conducted on the returned device.Visual analysis of the returned sample revealed that the el5ml device was returned with no apparent damage.In an attempt to replicate the reported incident, the instrument was tested for functionality.During the analysis, the device was cycled and it fed and formed 4 conforming clips. upon testing, the jaws open and close without any difficulties.In addition, the device locked out as intended. although no conclusion could be reached on the cause of the reported event, the instructions for use do contain the following caution: do not excessively twist or torque the instrument jaws when positioning or firing the instrument on a tubular structure or vessel.Excessive twisting or torquing may result in clip malformation.Do not insert the clip applier through a trocar if a clip is present in the jaws.This may result in clip malformation, dislodged clips, or damage to the instrument.If a clip is present in the jaws, fully squeeze the trigger against the handle, then fully release the trigger to release the clip from the jaws before inserting the device through the trocar.The event described could not be confirmed as the device performed without any difficulties noted.As part of ethicon¿s quality process, all devices are manufactured, inspected, and released to approved specifications.A manufacturing record evaluation was performed for the finished device batch and lot number, and no non-conformances were identified.
|