Catalog Number UNK XIENCE
Device Problem Break (1069)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 11/28/2023
Event Type  malfunction  
Event Description
It was reported that the procedure was to treat a lesion in an unspecified artery.An unknown xience stent delivery system (sds) had a fractured wire [delivery catheter].There was no reported adverse patient effect and there was no reported clinically significant delay in the procedure.No additional information was provided.
 
Manufacturer Narrative
D4: the udi is unknown because the part/lot number were not provided.Manufacturer's investigation is still pending at this time.Results and conclusions will be provided in the final report.
 
Manufacturer Narrative
The device was not returned for analysis.A review of the electronic lot history record (elhr) and lot level similar incident search for this product were not performed because the part and lot numbers were not reported and the product was not returned for analysis.The investigation was unable to determine a conclusive cause for the reported shaft break.It is possible that the device was inadvertently mishandled during advancement, as no failures were reported during preparation, resulting in the reported break.Without having the device to examine and limited information provided, a conclusive cause cannot be determined.There is no indication of a product quality issue with respect to manufacture, design or labeling.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
XIENCE EVEROLIMUS ELUTING CORONARY STENT SYSTEM
Type of Device
DRUG ELUTING CORONARY STENT DELIVERY SYSTEM
Manufacturer (Section D)
ABBOTT VASCULAR
26531 ynez rd.
temecula CA 92591 4628
Manufacturer (Section G)
ABBOTT VASCULAR, REG # 3005718570 (P099)
cashel road
clonmel tipperary
EI  
Manufacturer Contact
lindsey bell
26531 ynez rd.
temecula, CA 92591-4628
9519143996
MDR Report Key18368639
MDR Text Key331296509
Report Number2024168-2023-14282
Device Sequence Number1
Product Code NIQ
Combination Product (y/n)Y
Reporter Country CodeBR
PMA/PMN Number
P070015
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative,Distributor
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 01/16/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberUNK XIENCE
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 11/28/2023
Initial Date FDA Received12/20/2023
Supplement Dates Manufacturer Received12/28/2023
Supplement Dates FDA Received01/16/2024
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-