Catalog Number 367856 |
Device Problems
Crack (1135); Fluid/Blood Leak (1250)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 11/25/2023 |
Event Type
malfunction
|
Event Description
|
It was reported when using the bd vacutainer® k2 edta (k2e) 5.4mg blood collection tubes is cracked and leaked during blood collection.There was no report of impact to patient or user.
|
|
Manufacturer Narrative
|
H.3.A device evaluation is anticipated, but has not yet begun.Upon completion of the investigation, a supplemental report will be filed.
|
|
Manufacturer Narrative
|
H.6.Investigation summary: bd had not received samples or photos for investigation.Therefore, 100 retention samples from bd inventory were evaluated by visual examination and no issues were observed as all samples met specifications.Based on a review of the device history record for the incident lot, all product specifications and requirements for lot release were met.There were no related quality issues during manufacturing of the product.This complaint is unable to be confirmed for the indicated failure mode.Bd was not able to identify a root cause for the indicated failure mode.Complaints received for this device and reported condition will continue to be tracked and trended.Our business team regularly reviews the collected data for identification of emerging trends.
|
|
Event Description
|
It was reported when using the bd vacutainer® k2 edta (k2e) 5.4mg blood collection tubes is cracked and leaked during blood collection.There was no report of impact to patient or user.
|
|