It was reported that during an implantation procedure, the subcutaneous implantable cardioverter defibrillator (s-icd) system was an attempted implant due to the s-icd being damaged or defective.Subsequently, an new s-icd system was successfully implanted.No adverse patient effects were reported.
|
It was reported that during an implantation procedure, the subcutaneous implantable cardioverter defibrillator (s-icd) system was an attempted implant due to the s-icd being damaged or defective.Subsequently, an new s-icd system was successfully implanted.No adverse patient effects were reported.Additional information was received indicating that during the procedure, the pulse generator would not interrogate and there were tones could not be elicited.When the device eventually connected, there was no sensing signal.High, out-of-range shock impedance measurements of greater than 400 ohms were also observed.Eventually, they realized the programmer was connecting to a spare device in the case cart, and the attempted s-icd was dead.Subsequently, the new s-icd system was implanted.No adverse patient effects were reported.
|