Model Number N/A
Device Problem Physical Resistance/Sticking (4012)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 11/02/2023
Event Type  malfunction  
Event Description
It was reported that the ratchet handle does not move up or down.The event occurred during kit inspection.There was no reported patient harm, or delay.Due diligence is complete, no further information is available at this time.
 
Manufacturer Narrative
This event is recorded with zimmer biomet under (b)(4).This medwatch is being filed as an initial report.E1 phone:(b)(6).G2 foreign: japan.Customer has indicated that the product is in process of being returned to zimmer biomet for investigation.Once the investigation has been completed, a follow-up mdr will be submitted.
 
Event Description
No additional event information is available.
 
Manufacturer Narrative
This event has been recorded by zimmer biomet under (b)(4).This medwatch is being filed to relay additional information.Review of the most recent repair record determined that the ratchet handle will not attach to unit.Scratches on the main body roller shaft.Repaired roller shaft, clearance was adjusted and resolved the reported issue.Review of the device history record identified no deviations or anomalies during manufacturing related to the reported event.A definitive root cause cannot be determined.The event is confirmed.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ZIMMER SKIN GRAFT MESHER
Type of Device
EXPANDER, SURGICAL, SKIN GRAFT
Manufacturer (Section D)
ZIMMER SURGICAL, INC.
200 west ohio avenue
dover OH 44622
Manufacturer (Section G)
ZIMMER SURGICAL, INC.
200 west ohio avenue
dover OH 44622
Manufacturer Contact
jennifer rapsavage
56 e. bell drive
warsaw, IN 46582
5745260384
MDR Report Key18371182
MDR Text Key331080306
Report Number0001526350-2023-01645
Device Sequence Number1
Product Code GFD
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Distributor
Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date 04/23/2024
2 Devices were Involved in the Event: 1   2  
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberN/A
Device Catalogue Number00770100000
Device Lot NumberN/A
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/02/2023
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 11/28/2023
Initial Date FDA Received12/21/2023
Supplement Dates Manufacturer Received04/23/2024
Supplement Dates FDA Received04/23/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured06/02/1998
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Removal/Correction NumberN/A
Patient Sequence Number1
Patient SexPrefer Not To Disclose
Patient EthnicityNon Hispanic
-
-