MAUDE Adverse Event Report: STRYKER-ENDOSCOPY LAKELAND 5300; ELECTROSURGICAL, CUTTING & COAGULATION ACCESSORIES, LAPAROSCOPIC & ENDOSCOPIC,

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER-ENDOSCOPY LAKELAND 5300; ELECTROSURGICAL, CUTTING & COAGULATION ACCESSORIES, LAPAROSCOPIC & ENDOSCOPIC, Back to Search Results
Model Number LF1837
Device Problems Use of Device Problem (1670); Failure to Seal (4070)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 11/01/2023
Event Type  malfunction  
Manufacturer Narrative
Additional information will be provided once the investigation has been completed.
 
Event Description
It was reported the device was taking longer than normal to cauterize.No information was provided regarding patient injury, medical intervention, or extended procedure time reported.These are commonly used devices that are readily available.
 
Manufacturer Narrative
The device was not returned to stryker sustainability solutions for evaluation.As the complaint device was not returned, a conclusive root cause could not be determined and the reported event could not be confirmed.A review of the dhr for the reported lot/serial number supports that the device met all inspection and test criteria prior to release from stryker.The reported event could be attributed to: use error: thumb trigger button (activation button) not engaged throughout the entire seal cycle.Environmental disturbance: noise.Failure mode: active electrode short, electrical connections damaged.Failure mode: device not properly cleaned.Environmental disturbance: contamination of device in pre-op inspection.The instructions for use (ifu) state: sealer/divider without nano-coating (model lf1837) instructions for use (ifu): do not place the vessel and/or tissue in the jaw hinge.Place the vessel and/or tissue in the center of the jaws.Do not use this instrument on vessels larger than 7 mm in diameter.If the instrument shaft is visibly bent, discard and replace the instrument.A bent shaft may prevent the instrument from sealing or cutting properly.Eliminate tension on the tissue when sealing and cutting to ensure proper function.Use caution when grasping, manipulating, sealing, and dividing large tissue bundles.Do not attempt to seal or cut over clips or staples as incomplete seals will be formed.Contact between an active electrode and any metal objects may result in alternate site burns or incomplete seals.Do not activate the ligasure system until the lever has been latched.Activating the system before latching may result in improper sealing and may increase thermal spread to tissue outside of the surgical site.Prior to cutting the seal, inspect the vessel or tissue to ensure proper sealing.Keep the instrument jaws clean.Build-up of eschar may reduce the seal and/or cutting effectiveness.Wipe jaw surfaces and edges with a wet gauze pad as needed.If the generator provides multiple power settings, use the lowest power needed to achieve the intended effect.Do not overfill the jaws of the instrument with tissue, as this may reduce device performance.Use caution during surgical procedures in which patients exhibit certain types of vascular pathology (atherosclerosis, aneurysmal vessels, etc.).For best results, apply the seal to unaffected vasculature.Energy-based devices, such as es pencils or ultrasonic scalpels that are associated with thermal spread, should not be used to transect seals.The reported event will continue to be monitored through post-market surveillance.Should the device become available for return, the investigation will be reopened.H3 other text : 81.
 
Event Description
It was reported the device was taking longer than normal to cauterize.There was no patient injury or medical intervention and extended procedure time reported was a few minutes.These are commonly used devices that are readily available.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NA
Type of Device
ELECTROSURGICAL, CUTTING & COAGULATION ACCESSORIES, LAPAROSCOPIC & ENDOSCOPIC,
Manufacturer (Section D)
STRYKER-ENDOSCOPY LAKELAND 5300
5300 region ct
lakeland FL 33815
Manufacturer (Section G)
STRYKER-ENDOSCOPY LAKELAND 5300
5300 region ct
lakeland FL 33815
Manufacturer Contact
andy dobos
1810 w. drake drive
tempe, AZ 85283
8888883433
MDR Report Key18378561
MDR Text Key331374061
Report Number0001056128-2023-00031
Device Sequence Number1
Product Code NUJ
Combination Product (y/n)N
Reporter Country CodeCA
PMA/PMN Number
K172856
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 02/13/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/21/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberLF1837
Device Catalogue NumberLF1837RR
Device Lot Number15082558
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received01/17/2024
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured07/10/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? Yes
Type of Device Usage Reuse
Patient Sequence Number1
-
-