Model Number 865350 |
Device Problem
No Audible Alarm (1019)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 12/15/2023 |
Event Type
malfunction
|
Manufacturer Narrative
|
Philips is in the process of obtaining additional information and the complaint is still under investigation.A follow-up report will be submitted upon completion of the investigation.
|
|
Event Description
|
It was reported the device indicates a speaker malfunction inop with no sound.The device was in clinical use, but no adverse event to the patient or user was reported.
|
|
Manufacturer Narrative
|
The device was sent to a philips authorized repair facility (rft) for evaluation.Functional tests were performed and the device the speaker produced audible sound.The reported problem was not confirmed.The rft replaced the speaker per current process and the device remains at the customer site.
|
|