Model Number 16-02-85
Device Problem Microbial Contamination of Device (2303)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 11/28/2023
Event Type  malfunction  
Event Description
Livanova deutschland received a report that a heater-cooler system 3t device was found to be contaminated with mycobacterium chelonae both pre (31 cfu/ml) and post-disinfection procedure (12 cfu/ml).The laboratory report confirming the reported contamination has been provided to livanova.The unit has been taken out of service by the customer.There is no report of patient injury.
 
Manufacturer Narrative
A.1.-a.5.There was no patient involvement.G.5.The heater-cooler 16-02-80 is not distributed in the usa and it is similar to heater-cooler 16-02-85, which is distributed in the usa (510(k) number: k191402).H.9.Livanova deutschland implemented a field safety notice for disinfection and cleaning of heater-cooler devices.The z number is z-2076/2081-2015.H10: livanova deutschland manufactures the heater-cooler system 3t.The incident occurred in hudson, florida.Livanova initiated an investigation.If any additional information pertinent to the reported event is received, it will be provided in a supplemental report.
 
Manufacturer Narrative
Livanova was informed the 3t heater cooler was taken out of service.Through a subsequent follow up, livanova was informed the device is cleaned regularly according to the instructions for use and it is placed inside the operating room, with the fan positioned opposite to the patient at an estimated distance of about 2 meters from the surgery field.The h2o2 level is daily checked when the device is being used.When the device is not used, it is stored full of water and the h2o2 daily check is not executed during weekends.No patient reusable heating blankets are used by the hospital and the 3t aerosol collection set is replaced after the allowed use period.A disposable pall-aquasafe water filter with a 0.2 m membrane is used for tap water.The livanova device is used in combination with medtronic myeotherb and it is connected through opaque disposable tubing.Based on collected findings, it cannot be excluded that the above deviation from ifu's identified relating to the missed daily check of h2o2 during weekends when the device is stored full of water may have caused or contributed to reported contamination.Livanova has implemented a strategy to progressively decrease the probability of bacteria growth in the hc device by applying multiple measures implemented over the past few years through dedicated capa and a field action.The risk is in the acceptable region.No specific action was currently deemed necessary.Livanova will keep monitoring the market.
 
Event Description
See initial report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HEATER-COOLER SYSTEM 3T
Type of Device
CONTROLLER, TEMPERATURE, CARDIOPULMONARY BYPASS
Manufacturer (Section D)
LIVANOVA DEUTSCHLAND
lindberghstrasse 25
munich
Manufacturer (Section G)
LIVANOVA DEUTSCHLAND GMBH
lindberghstrasse 25
munich 80309
GM   80309
Manufacturer Contact
enrico greco
14401 w. 65th way
arvada, CO 80004
MDR Report Key18379915
MDR Text Key331383839
Report Number9611109-2023-00646
Device Sequence Number1
Product Code DWC
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional
Reporter Occupation Other Health Care Professional
Remedial Action Recall
Type of Report Initial,Followup
Report Date 01/30/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number16-02-85
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 11/28/2023
Initial Date FDA Received12/21/2023
Supplement Dates Manufacturer Received01/03/2024
Supplement Dates FDA Received01/31/2024
Was Device Evaluated by Manufacturer? No
Date Device Manufactured02/03/2011
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Removal/Correction NumberZ-2076/2081-2015