Reportable based on device analysis completed on 05dec2023.It was reported that the coil could not advance inside the catheter and got kinked after removed.The target lesion was located in the internal iliac artery.The direxion transend-14 system was selected for use.During the procedure, the catheter was not in a correct position as the physician was not satisfied on the initial placement of the coil.The physician withdrawn the device and after being flushed with heparin saline, it was deployed again, however, it was noted that the coil could not be advance inside the tail end of the catheter.The device was removed by following the contrast directly out of the body.When checked after removal, it was found that the microcatheter was kinked.The procedure completed with another of same device.No complications were reported, and patient was stable post procedure.However, device analysis revealed that the shaft was fractured 124.7cm from the hub.
|