Model Number N/A |
Device Problem
Physical Resistance/Sticking (4012)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 11/28/2023 |
Event Type
malfunction
|
Event Description
|
It was reported that at unknown timing an air dermatome was faulty.It was completely stuck.There was no report of harm or injury.Due diligence is in process.No new information has been received.No adverse events were reported as a result of this malfunction.
|
|
Manufacturer Narrative
|
This event has been recorded by zimmer biomet under (b)(4).G2 foreign: finland evaluation and investigation is in process.Once the investigation is complete, a supplemental report will be filed.
|
|
Event Description
|
No additional event information available.
|
|
Manufacturer Narrative
|
This event has been recorded by zimmer biomet under (b)(4).This medwatch is being filed to relay additional information.Review of the most recent repair record determined the motor speed was unstable and the lever was deformed.The lever, motor and sleeve were replaced and resolved the reported issue.Review of the device history record identified no deviations or anomalies during manufacturing related to the reported event.A definitive root cause cannot be determined.The event cannot be confirmed.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
|