Model Number CDHFA500Q
Device Problem Over-Sensing (1438)
Patient Problems Insufficient Information (4580); No Clinical Signs, Symptoms or Conditions (4582)
Event Date 11/26/2023
Event Type  malfunction  
Event Description
It was reported that the patient presented via a remote transmission showing episodes of non-sustained right ventricular oversensing (nsrvo).A review of the stored emg noted the over-sensing was due to post-paced t-wave oversensing (pptwos).Programming changes were not made at this time.
 
Event Description
New information received notes programming changes were made in clinic on (b)(6) 2024.The patient was doing well, and no complaints had been received since the changes.
 
Manufacturer Narrative
Section h6: codes updated/corrected to reflect patient condition and programming changes.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GALLANT HF
Type of Device
NO MATCH
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer Contact
richard williamson
15900 valley view court
sylmar, CA 91342
MDR Report Key18381266
MDR Text Key331178090
Report Number2017865-2023-95329
Device Sequence Number1
Product Code NIK
UDI-Device Identifier05415067032010
UDI-Public05415067032010
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P030054
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup
Report Date 02/13/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberCDHFA500Q
Device Lot NumberA000147927
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 11/27/2023
Initial Date FDA Received12/22/2023
Supplement Dates Manufacturer ReceivedNot provided
02/12/2024
Supplement Dates FDA Received02/13/2024
02/13/2024
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured09/24/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
OPTISURE; QUARTET; TENDRIL
Patient SexMale
-
-