Model Number 97715 |
Device Problems
Intermittent Continuity (1121); Failure to Deliver Energy (1211); Insufficient Information (3190)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 11/08/2023 |
Event Type
malfunction
|
Manufacturer Narrative
|
Continuation of d10: product id 97755 serial# (b)(6) product type recharger product id 97745 serial# (b)(6) product type programmer, patient b3: date is valid.G2.Foreign: canada medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
Information was received from a patient who was implanted with an implantable neurostimulator (ins) for unknown indications for use.It was reported that error code rm03 shows on the controller and p.77 in the bottom right.The therapy stops and it must be turned on again all the time.The patient did the reset steps by pressing the button and holding for about 20 seconds, has removed the lithium battery and placed it back but the issue persists.A replacement controller was requested, and the issue has resolved.No new event information.No new information.Additional information was received.It was reported that the implanted neurostimulator (ins) was not working properly, they were getting error code 77, system problem rm03, and had reset 4 times.Normal wear and tear was indicated.The patient received a new programmer a few weeks ago.Unknown if the issue has resolved.A replacement recharger kit was requested.
|
|
Search Alerts/Recalls
|
|
|