MAUDE Adverse Event Report: HILL-ROM BATESVILLE CENTRELLA MED-SURG; BED, AC-POWERED ADJUSTABLE HOSPITAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

HILL-ROM BATESVILLE CENTRELLA MED-SURG; BED, AC-POWERED ADJUSTABLE HOSPITAL Back to Search Results
Model Number P7900B100010
Device Problem Activation Problem (4042)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/13/2023
Event Type  malfunction  
Manufacturer Narrative
The baxter technician found the head actuator needed to be replaced.Per the hillrom service manual, it is necessary for the centrella¿ bed to have an effective maintenance program.We recommend that you do annual preventive maintenance.Make sure the cpr feature operates correctly.Repair or replace the siderail if necessary.A search of the baxter maintenance records did not show baxter performed any preventative maintenance on this bed.It is unknown if the facility performs preventative maintenance on their beds.The technician replaced the head actuator to resolve the reported event.Based on this information, no further action is required.
 
Event Description
Baxter received a report from a customer stating the cpr function was not working properly.The bed was located at the account.There was no patient/user injury reported.This report was filed in our complaint handling system as complaint #: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CENTRELLA MED-SURG
Type of Device
BED, AC-POWERED ADJUSTABLE HOSPITAL
Manufacturer (Section D)
HILL-ROM BATESVILLE
1069 state route 46 east
batesville IN 47006
Manufacturer Contact
daniela avila
1069 state route 46 east
batesville, IN 47006
8129310130
MDR Report Key18384426
MDR Text Key331249656
Report Number1824206-2023-01455
Device Sequence Number1
Product Code FNL
UDI-Device Identifier00887761985162
UDI-Public010088776198516211200327
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility,Company Representative
Reporter Occupation Other
Type of Report Initial
Report Date 12/22/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Model NumberP7900B100010
Was Device Available for Evaluation? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 12/13/2023
Initial Date FDA Received12/22/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/27/2020
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-