It was reported that the patient was presented for an implant procedure.During the procedure, after the implantable cardioverter defibrillator (icd) was connected to the lead, unknown pacing behavior was observed, where the ventricular pacing at a much shorter interval.Then, it was discovered that the sensitivity parameter had changed from the programmed parameter, and was reprogrammed to the nominal setting, which fixed the abnormal pacing behavior.Later, during post-operation interrogation, upon review of the parameters, it was found that the post-shock pacing setting was turned off, a change that was not intentionally made, and it was programmed back on.Finally, it was attempted to pair the icd with the monitoring application.After some difficulty, the application was paired to the device, but attempts to send a transmission failed as the application was unable to connect to the icd.The device was reinterrogated and it was observed that the bluetooth for remote monitoring was turned off and it was confirmed that the setting had not been touched.The setting was turned on, but subsequent attempts to connect the icd to the app and send a transmission failed.There were no patient consequences.
|
The reported event of phantom programming was not confirmed.Based on the information provided, no phantom programming occurred.The reported event of unable to pair with application was confirmed.Based on the information provided, it was determined that when the device was taken out of ship settings, the normal advertising was not set to 2 minutes, resulting in the connectivity issue.
|