SMITH & NEPHEW, INC. GII SPC P/S HOUS COLLET S 3-8; PROSTHESIS, KNEE, FEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, METAL/POLYMER
|
Back to Search Results |
|
Catalog Number 71440001 |
Device Problem
Detachment of Device or Device Component (2907)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 11/27/2023 |
Event Type
malfunction
|
Manufacturer Narrative
|
Internal complaint number: (b)(4).
|
|
Event Description
|
It was reported that, during a tka, the screw of one (1) gii spc p/s hous collet s 3-8 broke upon impaction of box osteotome.All pieces were removed by surgeon hand.The procedure was resumed, without any delay, using a s+n back-up device.Patient was not harmed as a consequence of this problem.As per the photos provided, the screw was detached from the device.
|
|
Manufacturer Narrative
|
E2: health professional, g3: report source.Section h3, h6: the associated device was returned and evaluated.The visual inspection revealed the collet washer, spring and collet thumbscrew is missing from one side of the device.These pieces were returned.The device shows signs of significant wear and use.A review of the production order did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.A review of complaint history revealed similar events for the listed device over the previous 12 months, but no similar events for the batch based on the historical data, this failure mode will be monitored for future complaints for any necessary corrective actions.A review of the risk management file revealed this failure mode was previously identified.The anticipated risk level is still adequate.A historical review concluded that there are no prior actions related to this product and event.At this time, we have no reason to suspect that the product failed to meet any specifications at the time of manufacture.This device is a reusable instrument that can be exposed to numerous surgeries.Damage from prolonged use, misuse or rough handling are likely potential factors that could contribute to the reported event.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.Based on this investigation, the need for corrective action is not indicated.Should additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.E1: name and address.
|
|
Search Alerts/Recalls
|
|
|