The reported events of failure to capture and failure to advance guidewire were not confirmed.As received, a complete lead was returned in one piece for analysis.Analysis found that the guidewire was able to be inserted and removed inside the lead without difficulties.Upon visual and x-ray examination, no anomalies were noted except for procedural damage.Upon electrical testing, no indication of conductor fractures or internal shorts were found.The s-curve height was measured within specification.
|