COOK INC NCIRCLE TIPLESS STONE EXTRACTOR; FFL DISLODGER, STONE, BASKET, URETERAL, METAL
|
Back to Search Results |
|
Model Number N/A |
Device Problem
Detachment of Device or Device Component (2907)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 12/06/2023 |
Event Type
malfunction
|
Event Description
|
It was reported that the wires detached from one side of the basket of a ncircle tipless stone extractor during a left ureteral stone extraction.A laser was used before basket made patient contact.The device was tested prior to use.It is unknown if the patient had tortuous, calcified, or scarred anatomy.Another device was used to complete the procedure.As reported, the patient did not experience any adverse effects or require any additional procedures due to this occurrence.
|
|
Manufacturer Narrative
|
Blank fields on this form indicate the information is unknown or unavailable.G4: pma/510(k) number = exempt; h3: device evaluated by mfg = other (81) - device evaluation has begun; however, a conclusion is not yet available.Device has been returned and preliminary evaluation has been performed, however, our investigation is ongoing and device evaluation summary will be included in our follow up report once our investigation has been completed.This report includes information known at this time.A follow-up report will be submitted should additional relevant information become available.This report is required by the fda under 21 cfr part 803 and is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement contained herein is intended to be an admission that any cook device is defective or malfunctioned, that a death or serious injury occurred, nor that any cook device caused, contributed to, or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
|
|
Manufacturer Narrative
|
Correction: h6 (annex a and annex g).Investigation ¿ evaluation: it was reported that the wires detached from one side of the basket of a ncircle tipless stone extractor during a left ureteral stone extraction.A laser was used before basket made patient contact.The device was tested prior to use.It is unknown if the patient had tortuous, calcified, or scarred anatomy.Another device was used to complete the procedure.The patient did not experience any adverse effects or require any additional procedures due to this occurrence.Reviews of the complaint history, device history record (dhr), instructions for use (ifu), manufacturing instructions, and quality control procedures, as well as a visual inspection of the returned device, were conducted during the investigation.Cook received one device for evaluation.The returned device was found to be nonfunctional due to the yellow support sheath having pulled free from the basket sheath.There was evidence of adhesive on the basket sheath, indicating proper assembly.Additionally, a document-based investigation evaluation was performed.In response to this incident, cook completed a review of the product device master record (dmr) and concluded that sufficient inspection activities are in place to identify this failure mode prior to distribution.Cook also reviewed the dhr for the reported lot and records no relevant non-conformances related to this incident.A database search for complaints on the reported lot found no additional complaints reported from the field.Cook concluded that no nonconforming product from this lot exists in house or in the field.Cook also reviewed product labeling.The product ifu, t _ ntse_ rev1; the ifu did not provide any information related to the reported issue.The information provided upon review of the dmr, dhr, ifu, and device evaluation suggests that the device was not manufactured out of specification, and that there are no nonconforming devices in house or out in the field.Based on the information provided, inspection of the returned device, and the results of the investigation, cook was unable to establish a cause for this event.The appropriate personnel have been notified.Cook will continue to monitor for similar complaints.Per the risk assessment no further action is required.This report is required by the fda under 21 cfr part 803 and is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement contained herein is intended to be an admission that any cook device is defective or malfunctioned, that a death or serious injury occurred, nor that any cook device caused, contributed to, or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
|
|
Event Description
|
No additional information regarding the patient and/or event has been received since the previous medwatch report was sent.
|
|