Model Number 10619 |
Device Problems
Break (1069); Material Deformation (2976)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 11/13/2023 |
Event Type
malfunction
|
Event Description
|
Reportable based on device analysis completed on (b)(6) 2023.It was reported that shaft kink occurred.During preparation of a 4.00 x 20mm synergy drug-eluting stent, the catheter was noted to be kinked and could not be delivered.The procedure was completed with another of the same device.There were no patient complications reported, and the patient was stable.However, returned device analysis revealed hypotube break.
|
|
Manufacturer Narrative
|
(b)(6).Device evaluated by mfr.: synergy ous mr 4.00 x 20mm was returned for analysis.A visual tactile and microscopic examination of the returned device confirmed that there was a break in the hypotube.The proximal section of the break, including the hub/manifold was not returned with the device.The break location measured 120cm proximal from the guidewire exit site.No kinks or damages were noted along the main section of the hypotube.A visual and tactile examination of the distal extrusion identified a break/dissection at 17.5cm distal to the guidewire exit site.The distal section of the break/dissection, including the balloon, crimped stent and tip was not returned with the device.A microscopic examination of the break site in the hypotube, identified no issues with the device which could have contributed to the break.A microscopic examination of the distal extrusion identified a break/dissection at 17.5cm distal to the guidewire exit site.No other damage was present.
|
|
Manufacturer Narrative
|
E1: initial reporter facility name: (b)(6) hospital.E1: initial reporter phone: (b)(6).Device evaluated by mfr.: synergy ous mr 4.00 x 20mm was returned for analysis.A visual and tactile examination identified a break in the hypotube.The proximal section of the break, including the hub/manifold was not returned with the device.The break location measured 120cm proximal from the guidewire exit site.No kinks or damages were noted along the main section of the hypotube.A visual and tactile examination of the distal extrusion identified a break/dissection at 17.5cm distal to the guidewire exit site.The distal section of the break/dissection, including the balloon, crimped stent and tip was not returned with the device.A microscopic examination of the break site in the hypotube, identified no issues with the device which could have contributed to the break.A microscopic examination of the distal extrusion identified a break/dissection at 17.5cm distal to the guidewire exit site.No other damage was present.
|
|
Event Description
|
Reportable based on device analysis completed on (b)(6) 2023.It was reported that shaft kink occurred.During preparation of a 4.00 x 20mm synergy drug-eluting stent, the catheter was noted to be kinked and could not be delivered.The procedure was completed with another of the same device.There were no patient complications reported, and the patient was stable.However, returned device analysis revealed hypotube break.It was further reported that the device was fractured right before deployment inside the patient and was removed with a snare.
|
|
Search Alerts/Recalls
|