Model Number 72404310 |
Device Problems
Mechanical Problem (1384); Material Puncture/Hole (1504); Adverse Event Without Identified Device or Use Problem (2993)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 11/28/2023 |
Event Type
Injury
|
Event Description
|
It was reported that the patient went to the hospital because this inflatable penile prosthesis (ipp) was not working properly.Two weeks later, a revision surgery was performed, during which a crack in the connecting tube of the left cylinder was identified.The pump was removed and replaced.There were no patient complications reported.
|
|
Event Description
|
It was reported that the patient went to the hospital because this inflatable penile prosthesis (ipp) was not working properly.Two weeks later, a revision surgery was performed, during which a crack in the connecting tube of the left cylinder was identified.The pump was removed and replaced.There were no patient complications reported.
|
|
Manufacturer Narrative
|
Upon receipt at our quality assurance laboratory, the returned component underwent a thorough analysis.The pump was visually and functionally inspected and tested for leaks.Visual inspection revealed that the tubing was cut down to the pump block; therefore, it was not returned for analysis.No leaks were identified in the pump; however, the component did not pass the activation test during functional evaluation.Based on the information available and analysis results, the pump not passing the functional examination could have caused or contributed to the reported clinical observation of mechanical issue.
|
|