MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION AMS INFLATABLE PENILE PROSTHESIS; DEVICE IMPOTENCE MECHANICAL/HYDRAULIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION AMS INFLATABLE PENILE PROSTHESIS; DEVICE IMPOTENCE MECHANICAL/HYDRAULIC Back to Search Results
Model Number 72404310
Device Problems Mechanical Problem (1384); Material Puncture/Hole (1504); Adverse Event Without Identified Device or Use Problem (2993)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 11/28/2023
Event Type  Injury  
Event Description
It was reported that the patient went to the hospital because this inflatable penile prosthesis (ipp) was not working properly.Two weeks later, a revision surgery was performed, during which a crack in the connecting tube of the left cylinder was identified.The pump was removed and replaced.There were no patient complications reported.
 
Event Description
It was reported that the patient went to the hospital because this inflatable penile prosthesis (ipp) was not working properly.Two weeks later, a revision surgery was performed, during which a crack in the connecting tube of the left cylinder was identified.The pump was removed and replaced.There were no patient complications reported.
 
Manufacturer Narrative
Upon receipt at our quality assurance laboratory, the returned component underwent a thorough analysis.The pump was visually and functionally inspected and tested for leaks.Visual inspection revealed that the tubing was cut down to the pump block; therefore, it was not returned for analysis.No leaks were identified in the pump; however, the component did not pass the activation test during functional evaluation.Based on the information available and analysis results, the pump not passing the functional examination could have caused or contributed to the reported clinical observation of mechanical issue.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AMS INFLATABLE PENILE PROSTHESIS
Type of Device
DEVICE IMPOTENCE MECHANICAL/HYDRAULIC
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer Contact
farshad fahimi
4100 hamline avenue north
building c
saint paul, MN 55112
MDR Report Key18395496
MDR Text Key331347915
Report Number2124215-2023-73811
Device Sequence Number1
Product Code FHW
UDI-Device Identifier00878953003986
UDI-Public00878953003986
Combination Product (y/n)N
Reporter Country CodeKS
PMA/PMN Number
N970012
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,Health Professional
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 03/27/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date04/18/2022
Device Model Number72404310
Device Catalogue Number72404310
Device Lot Number0172148003
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 12/13/2023
Initial Date FDA Received12/26/2023
Supplement Dates Manufacturer Received03/08/2024
Supplement Dates FDA Received03/28/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured04/18/2017
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
CYLINDERS: UPN: 72404273 / LOT: 164366003; CYLINDERS: UPN: 72404273 / LOT: 164366003; RESERVOIR: UPN: 720182-01 / LOT: 159394014; RESERVOIR: UPN: 720182-01 / LOT: 159394014
Patient Outcome(s) Hospitalization; Required Intervention;
Patient Age65 YR
Patient SexMale
-
-