ST. JUDE MEDICAL, INC.(CRM-SYLMAR) GALLANT HF; NO MATCH Back to Search Results
Model Number CDHFA500Q
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problem Wound Dehiscence (1154)
Event Date 12/07/2023
Event Type  Injury  
Event Description
It was reported that the asymptomatic patient presented for a follow-up in clinic with wound dehiscence on the incision site.The physician explanted the implantable cardioverter defibrillator (icd) and the leads to resolve the issue.The patient was in stable condition throughout the procedure.
 
Manufacturer Narrative
The product was returned.Interrogation and visual inspection were normal.Analysis found that the device was received with battery voltage above elective replacement indicator (eri) level.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GALLANT HF
Type of Device
NO MATCH
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer Contact
richard williamson
15900 valley view court
sylmar, CA 91342
MDR Report Key18396349
MDR Text Key331357913
Report Number2017865-2023-95608
Device Sequence Number1
Product Code NIK
UDI-Device Identifier05415067032010
UDI-Public05415067032010
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P030054
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date 02/16/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/27/2023
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Model NumberCDHFA500Q
Device Lot NumberA000143086
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received02/13/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured06/05/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
OPTISURE MRI LEAD; QUARTET LEADS; TENDRIL LEADS
Patient Outcome(s) Required Intervention;
Patient Age77 YR
Patient SexMale
-
-