MAUDE Adverse Event Report: ZIMMER SURGICAL, INC. ZIMMER AIR DERMATOME

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER SURGICAL, INC. ZIMMER AIR DERMATOME Back to Search Results
Model Number N/A
Device Problems Failure to Cut (2587); Patient Device Interaction Problem (4001)
Patient Problem Laceration(s) (1946)
Event Type  Injury  
Event Description
It was reported that during surgery an incomplete graft was produced, where there was holes and strips in the graft.There was harm associated and a 10 minute extension in procedure.Due diligence is complete.No additional information is available.No additional consequences have been reported as a result of this malfunction.
 
Manufacturer Narrative
This event has been recorded by zimmer biomet under (b)(4).Evaluation and investigation is in process.Once the investigation is complete, a supplemental report will be filed.
 
Manufacturer Narrative
This event has been recorded by zimmer biomet under cmp (b)(4).This medwatch is being filed to relay additional information.The following sections were updated/corrected.Review of the most recent repair record could not be completed because the device was not returned for evaluation and repair.Device history record (dhr) review was unable to be performed as the lot number of the device involved in the event is unknown.A definitive root cause cannot be determined.The event cannot be confirmed.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event Description
Due diligence is complete and no additional information is available.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ZIMMER AIR DERMATOME
Type of Device
DERMATOME
Manufacturer (Section D)
ZIMMER SURGICAL, INC.
200 west ohio avenue
dover OH 44622
Manufacturer (Section G)
ZIMMER SURGICAL, INC.
200 west ohio avenue
dover OH 44622
Manufacturer Contact
jennifer rapsavage
56 e. bell drive
warsaw, IN 46582
5745260384
MDR Report Key18397624
MDR Text Key331376310
Report Number0001526350-2023-01687
Device Sequence Number1
Product Code GFD
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 12/27/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/27/2023
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberN/A
Device Catalogue Number00880100100
Device Lot NumberN/A
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received11/28/2023
Was Device Evaluated by Manufacturer? No
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Removal/Correction NumberN/A
Patient Sequence Number1
Treatment
DERMATOME BLADESPN 00880000010LN UNKNOWN
Patient Outcome(s) Required Intervention;
Patient SexPrefer Not To Disclose
-
-