MED-EL ELEKTROMEDIZINISCHE GERAETE GMBH MED-EL MAESTRO COCHLEAR IMPLANT SYSTEM Back to Search Results
Model Number MI1200 SYNCHRONY PIN
Device Problems Impedance Problem (2950); Insufficient Information (3190)
Patient Problem Failure of Implant (1924)
Event Date 12/12/2023
Event Type  malfunction  
Event Description
The user's hearing performance with the device is affected after a trauma to the head.The user was reimplanted on (b)(6) 2023.
 
Manufacturer Narrative
The device has been explanted and should be returned to the manufacturer for evaluation.When available, a device failure analysis will be submitted as a follow up report.
 
Manufacturer Narrative
Conclusion: the investigation results revealed overload fractures in the active electrode confirming that the device has failed due to an external impact to the active electrode.All the problems given in the recipient report appear to match well with this finding.This is a final report.
 
Event Description
The user's hearing performance with the device is affected after a trauma to the head.The user was re-implanted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MED-EL MAESTRO COCHLEAR IMPLANT SYSTEM
Type of Device
COCHLEAR IMPLANT
Manufacturer (Section D)
MED-EL ELEKTROMEDIZINISCHE GERAETE GMBH
innsbruck
AU 
Manufacturer Contact
laura simonotti
fuerstenweg 77a
innsbruck 6020
AU   6020
57788
MDR Report Key18398071
MDR Text Key331385503
Report Number9710014-2023-01149
Device Sequence Number1
Product Code MCM
UDI-Device Identifier09008737094010
UDI-Public(01)09008737094010
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
P000025
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Study,Company Representative
Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date 02/20/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date03/08/2021
Device Model NumberMI1200 SYNCHRONY PIN
Was Device Available for Evaluation? Device Returned to Manufacturer
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 12/14/2023
Initial Date FDA Received12/27/2023
Supplement Dates Manufacturer Received12/14/2023
Supplement Dates FDA Received02/20/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/08/2018
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age55 YR
Patient SexFemale
-
-