Model Number MI1200 SYNCHRONY PIN |
Device Problems
Impedance Problem (2950); Insufficient Information (3190)
|
Patient Problem
Failure of Implant (1924)
|
Event Date 12/12/2023 |
Event Type
malfunction
|
Event Description
|
The user's hearing performance with the device is affected after a trauma to the head.The user was reimplanted on (b)(6) 2023.
|
|
Manufacturer Narrative
|
The device has been explanted and should be returned to the manufacturer for evaluation.When available, a device failure analysis will be submitted as a follow up report.
|
|
Manufacturer Narrative
|
Conclusion: the investigation results revealed overload fractures in the active electrode confirming that the device has failed due to an external impact to the active electrode.All the problems given in the recipient report appear to match well with this finding.This is a final report.
|
|
Event Description
|
The user's hearing performance with the device is affected after a trauma to the head.The user was re-implanted.
|
|