ZIMMER BIOMET, INC. PERSONA PARTIAL MEDIAL ARTICULAR SURFACE PROVISIONAL 8MM SIZE E; INSTRUMENT, KNEE Back to Search Results
Model Number N/A
Device Problem Fracture (1260)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/06/2023
Event Type  malfunction  
Manufacturer Narrative
(b)(4).The complainant has indicated that the product will not be returned to zimmer biomet for investigation.The investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.H3 other text : investigation incomplete.
 
Event Description
It was reported during a unicompartmental knee arthroplasty that the tibial articular surface provisional fractured upon removal.No adverse events were reported as a result of this malfunction.
 
Manufacturer Narrative
This follow-up report is being submitted to relay additional information.Updated: b4, g3, g6, h2, h3, h6, and h10.Complaint sample was evaluated and the reported event was confirmed.Visual examination of the returned product identified signs of repeated use and the device was found to be fractured.Device history record was reviewed and no discrepancies were found.Root cause was unable to be determined.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event Description
No further event information available at the time of this report.
 
Manufacturer Narrative
This follow-up report is being submitted to relay additional information.
 
Event Description
No further event information available at the time of this report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PERSONA PARTIAL MEDIAL ARTICULAR SURFACE PROVISIONAL 8MM SIZE E
Type of Device
INSTRUMENT, KNEE
Manufacturer (Section D)
ZIMMER BIOMET, INC.
56 e. bell drive
p.o. box 587
warsaw IN 46581
Manufacturer (Section G)
ZIMMER BIOMET, INC.
56 e. bell drive
p.o. box 587
warsaw IN 46581
Manufacturer Contact
jennifer rapsavage
56 e. bell dr.
warsaw, IN 46582
5745260384
MDR Report Key18400519
MDR Text Key331542373
Report Number0001825034-2023-03077
Device Sequence Number1
Product Code HSX
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K161592
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Distributor
Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date 03/01/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberN/A
Device Catalogue Number42518100508
Device Lot Number64460720
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 12/06/2023
Initial Date FDA Received12/27/2023
Supplement Dates Manufacturer Received01/08/2024
02/27/2024
Supplement Dates FDA Received02/05/2024
03/01/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured01/08/2020
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Removal/Correction NumberN/A
Patient Sequence Number1
Patient SexPrefer Not To Disclose
-
-