Model Number P300NJ
Device Problems No Device Output (1435); No Flow (2991)
Patient Problem Insufficient Information (4580)
Event Date 11/01/2023
Event Type  malfunction  
Manufacturer Narrative
Date of event is unknown; no information has been provided to date.One device was received.No physical damage was found on the device.Per functional testing, the reported event was not reproduced.The complaint was not confirmed.The product's history records were reviewed and there were no non-conformances nor service-related issues that would have resulted in the reported complaint.No action was taken since the complaint was not confirmed.
 
Event Description
It was reported that during patient use there was no ventilation.Adverse patient effects are unknown.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PNEUPAC PARAPAC PLUS, JAPANESE PANEL
Type of Device
VENTILATOR, EMERGENCY, POWERED (RESUSCITATOR)
Manufacturer (Section D)
SMITHS MEDICAL INTERNATIONAL LTD.
enterprise way
luton LU3 4 BU
UK  LU3 4BU
Manufacturer Contact
reed covert
6000 nathan lane north
minneapolis, MN 55442
2247062300
MDR Report Key18402540
MDR Text Key331731149
Report Number9611178-2023-00014
Device Sequence Number1
Product Code BTL
UDI-Device Identifier00(01)(17)(10)
UDI-Public(01)(17)(10)
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K123957
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Company Representative
Reporter Occupation Other
Type of Report Initial
Report Date 12/27/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/27/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberP300NJ
Device Catalogue NumberP300NJJP
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/27/2023
Date Manufacturer Received11/27/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured11/16/2017
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-