Model Number 0998-00-0800-53 |
Device Problem
Optical Problem (3001)
|
Patient Problems
Insufficient Information (4580); No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 12/19/2023 |
Event Type
malfunction
|
Manufacturer Narrative
|
A supplemental report will be submitted upon completion of our investigation.
|
|
Event Description
|
It was reported that, the cardiosave intra-aortic balloon pump (iabp) has a fiber optic sensor error.
|
|
Event Description
|
It was reported that during use on patient, the cardiosave intra-aortic balloon pump (iabp) has a fiber optic sensor error.There was no patient harm reported.
|
|
Manufacturer Narrative
|
Additional customer contact information: name: (b)(6), email: (b)(6), phone: (b)(6).A getinge field service engineer (fse) stated external visual inspection was done.All checked ok not able to reproduce failure.Able to verify te53 fos continuous bit fail in logs.Cleaned fiber optic functional testing and safety check to factory specifications.Returned to the customer.
|
|
Event Description
|
N/a.
|
|