It was reported that during a routine follow up, it was found that this left ventricular (lv) lead exhibited low out of range pacing impedance measurements, with intermittent jumps.Additionally, it was observed that the lv pacing threshold trend included considerably high measurements, but all other electrical measurements were found to be in range.A damaged lead electrode ring is suspected to be the cause of the reported observations.A patient follow up was scheduled.At this time, no further information is available.The lead remains in service and adverse patient effects have been reported.Additional information was received.The patient has not yet been re-evaluated.No adverse patient effects were reported.The lead remains in service.
|
It was reported that during a routine follow up, it was found that this left ventricular (lv) lead exhibited low out of range pacing impedance measurements, with intermittent jumps.Additionally, it was observed that the lv pacing threshold trend included considerably high measurements, but all other electrical measurements were found to be in range.A damaged lead electrode ring is suspected to be the cause of the reported observations.A patient follow up was scheduled.At this time, no further information is available.The lead remains in service and adverse patient effects have been reported.Additional information was received.The patient has not yet been re-evaluated.No adverse patient effects were reported.The lead remains in service.Additional information was received.The patient underwent surgery, and the implantable device was explanted and replaced.The physician made the decision not to replace this lead, as they did not find evidence of an integrity concern during the procedure.No lead revision was performed at physicians discretion.The lead remains implanted and in-service.
|