• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STERIS MEXICO, S. DE R.L. DE C.V. V-PRO MAX 2 STERILIZER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STERIS MEXICO, S. DE R.L. DE C.V. V-PRO MAX 2 STERILIZER Back to Search Results
Device Problem Use of Device Problem (1670)
Patient Problem Burn(s) (1757)
Event Date 12/01/2023
Event Type  malfunction  
Manufacturer Narrative
A steris service technician arrived onsite following the reported event to inspect the sterilizer and found the unit to be operating properly.No issues with the function or operation of the sterilizer were identified, and the sterilizer was returned to service.The v-pro max 2 sterilizer operator manual states (6-16), "steris recommends (in accordance with ansi/aami guidelines) wearing chemical-resistant gloves when using the sterilization unit." the operator manual further states (1-2), "danger - chemical injury hazard: any visible liquids in the chamber or on the load must be treated as concentrated hydrogen peroxide.Observe all hydrogen peroxide handling precautions.When handling hydrogen peroxide, wear appropriate personal protective equipment." additionally, user facility personnel should ensure all instruments are properly dry prior to placement in the v-pro max 2 sterilizer.The v-pro max 2 sterilizer operator manual, (a-1) states, "dry all items thoroughly.Ensure all moisture is removed from all internal parts (including lumens).If not, residual hydrogen peroxide may remain at cycle completion and/or a cycle abort occurs.Only dry items are to be placed in sterilization unit." in-service training will be offered on the importance of wearing proper ppe, specifically gloves, while operating their vpro max 2 sterilizers.A follow-up mdr will be submitted when additional information becomes available.
 
Event Description
The user facility reported via pmda report that an employee obtained a burn on their finger when handling a pouch after it was processed in their v-pro max 2 sterilizer.The employee washed their hands and returned to work.No report of procedure delay.
 
Manufacturer Narrative
The technician counseled user facility personnel on the importance of wearing proper ppe, specifically gloves, while operating their v-pro max 2 sterilizer.No additional issues have been reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
V-PRO MAX 2 STERILIZER
Type of Device
STERILIZER
Manufacturer (Section D)
STERIS MEXICO, S. DE R.L. DE C.V.
avenida avante 790
parque industrial guadalupe
guadalupe, 67190
MX  67190
Manufacturer (Section G)
STERIS MEXICO, S. DE R.L. DE C.V.
avenida avante 790
parque industrial guadalupe
guadalupe, 67190
MX   67190
Manufacturer Contact
daniel davy
5960 heisley road
mentor, OH 44060
MDR Report Key18409138
MDR Text Key331718211
Report Number3005899764-2023-00071
Device Sequence Number1
Product Code MLR
UDI-Device Identifier00724995158361
UDI-Public00724995158361
Combination Product (y/n)N
Reporter Country CodeJA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 12/28/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/28/2023
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Date Manufacturer Received12/04/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured01/29/2023
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-