Catalog Number 2C8546
Device Problem Device Contamination with Chemical or Other Material (2944)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/05/2023
Event Type  malfunction  
Event Description
It was reported that a clearlink continu-flo solution set had dark particulate matter inside the tubing; further described as "something in the line." this occurred prior to patient use.There was no patient involvement.No additional information is available.
 
Manufacturer Narrative
G1: device manufacturer address 1: 600 mts.Oeste de entrada, principal ave.Las americas, parque industrial should additional relevant information become available, a supplemental report will be submitted.
 
Manufacturer Narrative
H10: the actual device was not available; however, a photograph of the sample was provided for evaluation.A visual inspection of the photo with the unaided eye revealed a brown embedded particle in the tubing which resembled burned resin from the extrusion process.Due to the nature of the sample no further testing could be performed.The reported condition was verified.The cause of the condition was determined to be a manufacturing related issue; probably due to burned resin which can be generated in the cannon and accumulate on the pin and bushing combinations.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CLEARLINK CONTINU-FLO SOLUTION SET
Type of Device
SET, ADMINISTRATION, INTRAVASCULAR
Manufacturer (Section D)
BAXTER HEALTHCARE CORPORATION
deerfield IL
Manufacturer (Section G)
BAXTER HEALTHCARE - CARTAGO
see h10
see h10
cartago 30106
CS   30106
Manufacturer Contact
25212 w. illinois route 120
round lake, IL 60073
2242702068
MDR Report Key18409807
MDR Text Key331508089
Report Number1416980-2023-06793
Device Sequence Number1
Product Code FPA
UDI-Device Identifier00085412048987
UDI-Public(01)00085412048987
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K203609
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Biomedical Engineer
Type of Report Initial,Followup
Report Date 02/02/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/28/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Catalogue Number2C8546
Device Lot NumberR23F25024
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received01/29/2024
Was Device Evaluated by Manufacturer? No
Date Device Manufactured06/27/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-