• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY ORTHOPAEDICS INC US QUICKSET 1PC FLEX DRILL BIT 25; DRILLS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY ORTHOPAEDICS INC US QUICKSET 1PC FLEX DRILL BIT 25; DRILLS Back to Search Results
Catalog Number 227425500
Device Problems Break (1069); Material Deformation (2976)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/20/2023
Event Type  malfunction  
Event Description
It was reported that drill bits became bent or broken during use.Surgical delay unknown.
 
Manufacturer Narrative
Product complaint #(b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803 (and/or part 4, as applicable).This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Manufacturer Narrative
Product complaint (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information, which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.If information is obtained, that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Manufacturer Narrative
Product complaint # (b)(4).Investigation summary: it was reported that drill bits became bent or broken during use.Surgical delay unknown.The device associated with this report was returned to depuy synthes for evaluation.Visual analysis of the returned sample revealed that quickset 1pc flex drill bit 25 was found bent at the drill bit tip.Additionally, the device shows an overall worn appearance consistent with normal and repetitive use.Broken condition was not observed.No other issues were identified.This condition is consistent with multiple uses under extreme eccentric loading resulting in premature bending or failure of the drill bit, however taking in consideration the age of the device (insert device age if applicable) the potential cause can be attributed to end of life.The lifecycle requirements of the device are event related and depend on the use and inspection of the device in clinical practice.As the device can be damaged on the first or 100th use, the device must be properly inspected prior to each surgical use.Refer to the device/country specific ifu for information related to end of life, reprocessing instructions, and inspection procedures.The overall complaint was confirmed as the observed condition of the quickset 1pc flex drill bit 25 would contribute to the complained device issue.Based on the investigation findings, the potential cause is traced to end of life, and it has been determined that no corrective and/or preventative action is required.As part of depuy synthes quality process, all devices are manufactured, inspected, and released to approved specifications.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Device history lot: the product investigation found no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).A manufacturing records evaluation (mre) was not performed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
QUICKSET 1PC FLEX DRILL BIT 25
Type of Device
DRILLS
Manufacturer (Section D)
DEPUY ORTHOPAEDICS INC US
700 orthopaedic drive
warsaw IN 46581 0988
Manufacturer (Section G)
DEPUY ORTHOPAEDICS, INC. 1818910
700 orthopaedic dr.
warsaw IN 46581 0988
Manufacturer Contact
kate karberg
700 orthopaedic dr.
warsaw, IN 46581
3035526892
MDR Report Key18409920
MDR Text Key331509906
Report Number1818910-2023-26123
Device Sequence Number1
Product Code HTW
UDI-Device Identifier10603295109129
UDI-Public10603295109129
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup
Report Date 12/28/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number227425500
Device Lot NumberPG333909
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 12/21/2023
Initial Date FDA Received12/28/2023
Supplement Dates Manufacturer Received01/25/2024
01/26/2024
Supplement Dates FDA Received01/25/2024
01/28/2024
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-