• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ABBOTT MEDICAL MITRACLIP¿ G4 STEERABLE GUIDE CATHETER (MDR); CATHETER, STEERABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT MEDICAL MITRACLIP¿ G4 STEERABLE GUIDE CATHETER (MDR); CATHETER, STEERABLE Back to Search Results
Catalog Number SGC0705
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problems Low Blood Pressure/ Hypotension (1914); Perforation (2001); Cardiac Tamponade (2226)
Event Date 12/08/2023
Event Type  Injury  
Manufacturer Narrative
Investigation is not yet complete.A follow-up report will be submitted with all additional relevant information.
 
Event Description
It was reported that a patient presented with grade 4 functional mitral regurgitation (mr) for a mitraclip procedure.During transseptal access, the blood pressure dropped to 60mmhg and a cardiac tamponade occurred.Surgical drainage was performed.The entry was from the right atrium, and it the cardiac tamponade was treated.The procedure continued after the patient was stabilized.One clip was placed on anterior 2 and posterior 2 leaflets.The mr was reduced to grade <1.After withdrawing the steerable guide catheter (sgc) a hole was found at the fossa ovalis.No additional information was provided.
 
Manufacturer Narrative
The device was not returned for evaluation.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot.Based on the information reviewed, the reported perforation appears to be related to procedural conditions.Perforation is listed in the instructions for use (ifu) as a known possible complication associated with mitraclip procedures.Serious injury/illness/impairment was a result of case-specific circumstances.There is no indication of a product issue with respect to manufacture, design or labeling.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MITRACLIP¿ G4 STEERABLE GUIDE CATHETER (MDR)
Type of Device
CATHETER, STEERABLE
Manufacturer (Section D)
ABBOTT MEDICAL
5050 nathan lane n
plymouth MN 55442
Manufacturer (Section G)
ABBOTT VASCULAR, REG # 3005070406
3885 bohannon drive
menlo park CA 94025
Manufacturer Contact
karen krouse
5050 nathan lane n
plymouth, MN 55442
6517565400
MDR Report Key18411218
MDR Text Key331526248
Report Number2135147-2023-05761
Device Sequence Number1
Product Code DRA
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K190167
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 01/18/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/28/2023
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Catalogue NumberSGC0705
Device Lot Number30828R1078
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received01/16/2024
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured08/28/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
MITRACLIP
Patient Outcome(s) Required Intervention; Other;
Patient Age84 YR
Patient SexFemale
Patient Weight40 KG
-
-