It was reported that this cardiac resynchronization therapy defibrillator (crt-d) system recorded a high out of range shock impedance measurement.Boston scientific technical services (ts) reviewed the device data and confirmed the out of range measurement.It was also discussed that all lead trends are showing initial variability and then stabilization.Further review of the crt-d system was recommended, and troubleshooting options were provided.The patient attended an in-clinic follow-up.The physician identified that the crt-d appears to be low in position and have some adipose tissue below the device, energy distribution was likely impacted by this reason.Induction testing was performed in-clinic.In the first attempt there was no ventricular fibrillation (vf) induction, on the second attempt the vf was induced and the shock delivered, however the shock was not effective and the vf terminated two seconds after the shock delivery.On the third attempt with reverse shock polarity the shock was not effective and an external shock was required to return the patient back to normal sinus rhythm.The shock impedance alert was increased to 150 ohms.Ts discussed possible options with the health care professional (hcp), however, there is no further action planned at this time.The crt-d system remains in service.No additional adverse patient effects were reported.
|