• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BIOSENSE WEBSTER INC QDOT MICRO¿ CATHETER; CATHETER, PERCUTANEOUS, CARDIAC ABLATION, FOR TREATMENT OF ATRIAL FIBRILLATION

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BIOSENSE WEBSTER INC QDOT MICRO¿ CATHETER; CATHETER, PERCUTANEOUS, CARDIAC ABLATION, FOR TREATMENT OF ATRIAL FIBRILLATION Back to Search Results
Catalog Number D139505
Device Problem Coagulation in Device or Device Ingredient (1096)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 11/30/2023
Event Type  malfunction  
Manufacturer Narrative
If additional information is received regarding this event, a supplemental 3500a report will be submitted to the fda.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by biosense webster, inc., or its employees that the report constitutes an admission that the product, biosense webster, inc., or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.Manufacturer's reference number: (b)(4).
 
Event Description
It was reported that a patient underwent an atrial fibrillation (afib) ablation procedure with a qdot micro¿ catheter and there was a clot that had formed.It was reported that when the physician was exchanging sheaths, the qdot micro¿ catheter was drawn back where it was noticed on the qdot tip, in-between electrodes two and three, there was a clot that had formed.The catheter was replaced, and the case continued.No adverse patient consequence was reported.
 
Manufacturer Narrative
It was reported that a patient underwent an atrial fibrillation (afib) ablation procedure with a qdot micro¿ catheter and there was a clot that had formed.It was reported that when the physician was exchanging sheaths, the qdot micro¿ catheter was drawn back where it was noticed on the qdot tip, in-between electrodes two and three, there was a clot that had formed.The catheter was replaced, and the case continued.No adverse patient consequence was reported.Device evaluation details: the device was returned to biosense webster (bwi) for evaluation.A visual inspection, temperature, impedance and patency tests of the returned device were performed in accordance with bwi procedures.Visual analysis of the returned device revealed reddish material inside the pebax and a hole in its surface.Temperature and impedance test was performed, and the device was found working correctly.No temperature or impedance issues were observed.A patency test was performed, and the device was flushing correctly.No obstructed holes were observed.No irrigation issues were observed.Based on the findings, the pebax issue may be related to the event described by the customer.A manufacturing record evaluation was performed for the finished device 31177540l number, and no internal action related to the complaint was found during the review.The issue reported by the customer was confirmed.It should be noted that product failure is multifactorial.The evaluation determined that char is a physical phenomenon of radiofrequency (rf).It can be the normal result of the ablation process.The instructions for use contain the following guideline: monitoring the temperature from the electrode during the application of rf current ensures that the irrigation flow rate is being maintained.Using tip temperature to guide ablation could result in deeper lesions and increased risk for collateral damage.Do not use the catheter without irrigation flow.Monitor the catheter tip temperature throughout the procedure to ensure adequate irrigation.If the temperature increases to 40°c during rf application, power delivery should be interrupted.Recheck the irrigation system prior to restarting rf application.The root cause for the pebax condition may be related to the handling of the device outside the manufacturing facilities; however, this could not be conclusively determined.As part of biosense webster's quality process, all devices are manufactured, inspected, and released to approved specifications.This product issue will be addressed through bwi's quality system.If additional information is received regarding this event, a supplemental 3500a report will be submitted to the fda.Manufacturer's reference number: (b)(4).
 
Manufacturer Narrative
On(b)(6)2024, the product investigation was completed as the complaint device was not returned.Since no device has been received for analysis, no product investigation can be performed, and the customer complaint cannot be confirmed.A manufacturing record evaluation (mre) was performed for the finished device 31177540l number, and no internal action related to the complaint was found during the review.Based on the completed mre, the manufactured date and expiration date have been provided.Therefore, fields d4.Expiration date and h4.Device manufacture date have been populated with appropriate information.If additional information is received regarding this event, a supplemental 3500a report will be submitted to the fda.Manufacturer's reference number: pc-(b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
QDOT MICRO¿ CATHETER
Type of Device
CATHETER, PERCUTANEOUS, CARDIAC ABLATION, FOR TREATMENT OF ATRIAL FIBRILLATION
Manufacturer (Section D)
BIOSENSE WEBSTER INC
31 technology drive, suite 200
irvine CA 92618
Manufacturer (Section G)
BIOSENSE WEBSTER INC (JUAREZ)
circuito interior norte
1820parque industrial salvacar
juarez 32599
MX   32599
Manufacturer Contact
kate karberg
31 technology dr
irvine, CA 92618
3035526892
MDR Report Key18414241
MDR Text Key331607119
Report Number2029046-2023-03109
Device Sequence Number1
Product Code OAE
UDI-Device Identifier10846835016758
UDI-Public10846835016758
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P210027
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Other
Type of Report Initial,Followup,Followup
Report Date 03/18/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberD139505
Device Lot Number31177540L
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 11/30/2023
Initial Date FDA Received12/28/2023
Supplement Dates Manufacturer Received01/05/2024
02/22/2024
Supplement Dates FDA Received01/29/2024
03/19/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured11/02/2023
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
UNK SHEATH
-
-