• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: RESMED PTY LTD ASTRAL 150 - AMER; VENTILATOR, CONTINUOUS (FACILITY/HOME)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

RESMED PTY LTD ASTRAL 150 - AMER; VENTILATOR, CONTINUOUS (FACILITY/HOME) Back to Search Results
Model Number 27003
Device Problem Protective Measures Problem (3015)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/04/2023
Event Type  malfunction  
Event Description
During evaluation by a third party service center, an astral device displayed an error message (sf179) related to the super capacitor.There was no patient harm or serious injury reported as a result of this incident.
 
Manufacturer Narrative
Visual inspection of the main circuit board revealed a leaky super capacitor.The main circuit board was replaced to address the reported complaint.The device was serviced and tested before it was returned to the customer.Based on all available evidence and complaint investigations of a similar nature, an investigation determined that the reported complaint was due to super capacitor electrolyte leak.Resmed's risk analysis for this failure mode concludes that the risk is acceptable.Resmed reference#: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ASTRAL 150 - AMER
Type of Device
VENTILATOR, CONTINUOUS (FACILITY/HOME)
Manufacturer (Section D)
RESMED PTY LTD
1 elizabeth macarthur drive
bella vista
sydney, nsw 2153
AS  2153
Manufacturer (Section G)
RESMED PTY LTD
1 elizabeth macarthur drive
bella vista
sydney, nsw 2153
AS   2153
Manufacturer Contact
jaklin aziz
1 elizabeth macarthur drive
bella vista
sydney, nsw 2153
AS   2153
MDR Report Key18414795
MDR Text Key331705999
Report Number3004604967-2023-00706
Device Sequence Number1
Product Code NOU
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K172875
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Reporter Occupation 505
Type of Report Initial
Report Date 12/28/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model Number27003
Device Catalogue Number27003
Was Device Available for Evaluation? No
Distributor Facility Aware Date12/04/2023
Initial Date Manufacturer Received 12/04/2023
Initial Date FDA Received12/28/2023
Date Device Manufactured05/05/2015
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage A
Patient Sequence Number1
-
-