Model Number CDDRA500Q
Device Problems Incorrect Interpretation of Signal (1543); Inappropriate/Inadequate Shock/Stimulation (1574)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/09/2023
Event Type  Injury  
Event Description
It was reported that the patient presented in clinic for a follow up.Device interrogation revealed inappropriate shock on the implantable cardioverter defibrillator (icd).No changes or intervention were reported.The patient condition was stable.
 
Manufacturer Narrative
Correction: missing h6 coding for incorrect interpretation of signal.
 
Event Description
New information noted that fast super ventricular tachycardia caused incorrect interpretation of signal was the cause of the inappropriate shock.
 
Manufacturer Narrative
Correction: new information noted that incorrect interpretation of signal was the cause of the inappropriate shock.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GALLANT DR
Type of Device
NO MATCH
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer Contact
richard williamson
15900 valley view court
sylmar, CA 91342
MDR Report Key18415865
MDR Text Key331576751
Report Number2017865-2023-95817
Device Sequence Number1
Product Code NIK
UDI-Device Identifier05415067032003
UDI-Public05415067032003
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P030054
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility,Company Representative
Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup,Followup
Report Date 01/09/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/28/2023
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date01/31/2021
Device Model NumberCDDRA500Q
Device Lot NumberS000077314
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Date Manufacturer Received01/08/2024
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured07/12/2020
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
2088TC/46.; 7120Q/58.
Patient Outcome(s) Other;
Patient Age62 YR
Patient SexFemale
-
-