DEPUY ORTHOPAEDICS INC US SMARTSET GHV GENTAMICIN 40G; BONE CEMENT : BONE CEMENT Back to Search Results
Catalog Number 3095040
Device Problems Difficult to Open or Remove Packaging Material (2922); Device Contaminated During Manufacture or Shipping (2969)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/20/2023
Event Type  malfunction  
Event Description
The nurse tried to tear the cement packet, but the gap between the outer packet and the sterile packet was very close, causing the sterile packet to tear as well.Therefore, the product is contaminated and cannot be used.Was surgery delayed due to the reported event? no, action taken when event occurred? open and use the new one, was procedure successfully completed? yes, were fragments generated? no, if yes, were they removed easily without additional intervention? unknown, patient status/ outcome / consequences no, was other medical intervention (e.G.X-rays, additional procedures, prescriptions, otc, revision) required: unknown, is the patient part of a clinical study unknown, (b)(4) device property of none, device in possession of none by checking this box i certify that all information that are known/available has been disclosed.If any new information will be made available, the additional information will be submitted through cst.True.
 
Manufacturer Narrative
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803 (and/or part 4, as applicable).This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.Dmf# - 13704 trade name gentamicin sulphate active ingredient(s) gentamicin sulphate dosage form - powder strength 1.0g active in our cements.
 
Event Description
Additional information received: a.Storage temperature before using is between 20-25 degrees.B.Or temperature during using cement is between 17-22 degrees.
 
Manufacturer Narrative
Product complaint # : (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Manufacturer Narrative
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Manufacturer Narrative
Product complaint (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information, that has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.If information is obtained, that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Manufacturer Narrative
Product complaint (b)(4).Investigation summary: the nurse tried to tear the cement packet, but the gap between the outer packet and the sterile packet was very close, causing the sterile packet to tear as well.Therefore, the product is contaminated and cannot be used.The product was not returned to depuy synthes, however photos were provided for review.See attachment img_1792.Jpeg, img_1791.Jpeg and img_1790.Jpeg.Visual inspection of the photos revealed that the smartset ghv gentamicin 40g was found with the carboard packaging open.The package contains two (2) pouches that are not sterile, both were found tear open, therefore the sterility of the bag that contains the cement was compromised.It is worth mentioning that the secondary packaging appears to be trapped in the foil seal.Dimensional inspection and document/specification review were not completed.The overall complaint was confirmed as the observed condition of the smartset ghv gentamicin 40g would contribute to the complained device issue.Based on the information currently available, a product issue was identified during the investigation of the sample received.This product issue will be addressed through depuy synthes quality system.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Device history lot: an nr was created to address current complaint condition.
 
Manufacturer Narrative
Product complaint # (b)(4).Investigation summary : according to the information received, the nurse tried to tear the cement packet, but the gap between the outer packet and the sterile packet was very close, causing the sterile packet to tear as well.Therefore, the product is contaminated and cannot be used.The device associated with this report was returned to depuy synthes for evaluation.Visual inspection of the returned sample revealed that the smartset ghv gentamicin 40g was found with the carboard packaging open.The package contains two (2) pouches that are not sterile, both were found tear open, therefore the sterility of the bag that contains the cement was compromised.It is worth mentioning that the secondary packaging appears to be trapped in the foil seal.A functional test was not performed as it is not applicable to the complaint condition.A dimensional inspection was not performed as it is not applicable to the complaint condition.The overall complaint was confirmed as the observed condition of the smartset ghv gentamicin 40g would contribute to the complained device issue.Based on the information currently available, a product issue was identified during the investigation of the sample received.This product issue will be addressed through depuy synthes quality system.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Device history lot : an nr was created to address current complaint condition.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SMARTSET GHV GENTAMICIN 40G
Type of Device
BONE CEMENT : BONE CEMENT
Manufacturer (Section D)
DEPUY ORTHOPAEDICS INC US
700 orthopaedic drive
warsaw IN 46581 0988
Manufacturer (Section G)
DEPUY CMW - 9610921
cornford rd
blackpool FY4 4 QQ
UK   FY4 4QQ
Manufacturer Contact
kate karberg
700 orthopaedic dr.
warsaw, IN 46581
3035526892
MDR Report Key18415950
MDR Text Key331705727
Report Number1818910-2023-26100
Device Sequence Number1
Product Code MBB
Combination Product (y/n)Y
Reporter Country CodeTH
PMA/PMN Number
OUS - SIMILA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup,Followup,Followup
Report Date 12/28/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number3095040
Device Lot Number4047299
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 12/19/2023
Initial Date FDA Received12/28/2023
Supplement Dates Manufacturer Received12/26/2023
01/18/2024
03/11/2024
03/15/2024
03/15/2024
Supplement Dates FDA Received01/08/2024
01/22/2024
03/14/2024
03/15/2024
03/15/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured01/23/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-