H10: the device was received for evaluation.A visual inspection with the naked eye was performed which observed that the check valve was assembled in the wrong orientation.A functional flow test was performed starting from where the spike was inserted into a saline bag; no leak was observed, however, the flow of liquid was only until the check valve and no flow was present after this location.The reported condition of leak was not verified; however, the device did not meet specification due to the incorrect orientation of the check valve.The cause of the condition was determined to be a manual assembly issue during the manufacturing process.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
|