• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DATASCOPE CORP. - MAHWAH CARDIOSAVE HYBRID, TYPE B PLUG; SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DATASCOPE CORP. - MAHWAH CARDIOSAVE HYBRID, TYPE B PLUG; SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL Back to Search Results
Model Number 0998-00-0800-53
Device Problem Failure to Power Up (1476)
Patient Problems Insufficient Information (4580); No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/20/2023
Event Type  malfunction  
Manufacturer Narrative
A supplemental report will be submitted upon completion of our investigation.Due to character restrictions in block e1 initial reporter : (b)(6).
 
Event Description
It was reported that, the cardiosave intra-aortic balloon pump (iabp) unit system will not boot up.
 
Manufacturer Narrative
Updated fields: h6 (type of investigation, investigation findings, component codes and investigation conclusions).Corrected fields: h6 (health effect ¿ clinical & impact code).A getinge field service engineer (fse) investigated the customer's complaint confirmed unit will not boot up in clinical mode, when unit was turned on just has a spinning wheel.Fse was able to get into the maintenance menu but there were no faults or error codes.Replaced pcba, video generator.Reassembled the unit and turned unit on but it needed the software updated , after update unit started up as it should , ran unit thru a complete calibration and electrical safety tests and all meets factory specifications.Ran unit on test balloon with no issues,unit was returned to customer.
 
Event Description
It was reported that during start up of unit, the cardiosave intra-aortic balloon pump (iabp) unit system will not boot up.There was no patient involvement.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CARDIOSAVE HYBRID, TYPE B PLUG
Type of Device
SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL
Manufacturer (Section D)
DATASCOPE CORP. - MAHWAH
1300 macarthur blvd
mahwah NJ
Manufacturer (Section G)
DATASCOPE CORP. - MAHWAH
1300 macarthur blvd
mahwah NJ
Manufacturer Contact
arelean guzman
1300 macarthur blvd
mahwah, NJ 
MDR Report Key18419805
MDR Text Key331622526
Report Number2249723-2023-05529
Device Sequence Number1
Product Code DSP
UDI-Device Identifier10607567108391
UDI-Public10607567108391
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K112372
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility,Company Representative
Reporter Occupation Biomedical Engineer
Type of Report Initial,Followup
Report Date 05/03/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number0998-00-0800-53
Device Catalogue Number0998-00-0800-53
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 12/20/2023
Initial Date FDA Received12/29/2023
Supplement Dates Manufacturer Received05/03/2024
Supplement Dates FDA Received05/03/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured05/23/2022
Is the Device Single Use? No
Type of Device Usage Reuse
Patient Sequence Number1
Treatment
UNKNOWN.
-
-