• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BAXTER HEALTHCARE CORPORATION FOLFUSOR; PUMP, INFUSION, ELASTOMERIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BAXTER HEALTHCARE CORPORATION FOLFUSOR; PUMP, INFUSION, ELASTOMERIC Back to Search Results
Catalog Number 2C4009K
Device Problem No Flow (2991)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 07/13/2023
Event Type  malfunction  
Manufacturer Narrative
E1: initial reporter address: (b)(6).Should additional relevant information become available, a supplemental report will be submitted.
 
Event Description
It was reported that a large volume folfusor had no flow.This occurred during patient infusion.The balloon remained completely full 48 hours after installation of the device.To resolve the event, the device was prepared again and reconnected.There was no report of patient injury or medical intervention associated with this event.No additional information is available.
 
Manufacturer Narrative
H4: the lot was manufactured october 04, 2022 to october 05, 2022.H10: one actual device was received for evaluation.The unit contained 229ml of fluid in the bladder.A visual inspection on the unit via the naked eye showed no evidence of fluid coming out at the distal end of the flow restrictor.A microscopic examination on the flow restrictor revealed the cause of the flow problem was due to a series of micro-air bubbles blocking the fluid path inside the lumen of the capillary glass.The capillary glass is located inside the flow restrictor housing.The reported condition was verified.The cause of the condition was determined to be user related as air bubbles inside the lumen of the capillary glass can be attributed to improper filling technique during product use (filling step).The product¿s instructions for use details the proper filling technique to avoid this type of occurrence.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FOLFUSOR
Type of Device
PUMP, INFUSION, ELASTOMERIC
Manufacturer (Section D)
BAXTER HEALTHCARE CORPORATION
deerfield IL
Manufacturer (Section G)
BAXTER HEALTHCARE - IRVINE
17511 armstrong avenue
building 3
irvine CA 92614
Manufacturer Contact
25212 w. illinois route 120
round lake, IL 60073
2242702068
MDR Report Key18419841
MDR Text Key331623359
Report Number1416980-2023-06837
Device Sequence Number1
Product Code MEB
UDI-Device Identifier00085412080130
UDI-Public(01)00085412080130
Combination Product (y/n)N
Reporter Country CodeFR
PMA/PMN Number
NA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Pharmacist
Type of Report Initial,Followup
Report Date 01/23/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Catalogue Number2C4009K
Device Lot Number22K003
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 12/05/2023
Initial Date FDA Received12/29/2023
Supplement Dates Manufacturer Received01/07/2024
Supplement Dates FDA Received01/23/2024
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age60 YR
Patient SexFemale
-
-