Catalog Number 14001-31
Device Problem Fluid/Blood Leak (1250)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 11/01/2023
Event Type  malfunction  
Manufacturer Narrative
Although the device was requested to be returned for evaluation, it has not been received.Without the returned device, a probable cause is unable to be determined.
 
Event Description
It was reported, on an unknown date in november, a primary plum set transparent disposable generated a leak during usage in a patient.When initiating the infusion, it was noted that the set was dripping/leaking.It was not reported that the set was contaminated, presented any biological risk, or contained any chemotherapeutic agent and that medicines were used with the set.The set was not reprocessed or re-sterilized.There was a delay in therapy, however, no one was harmed, and was not reported that an adverse event occurred as a result of this event.This is report six of eight.
 
Manufacturer Narrative
A video was provided by the customer.There was leakage observed from the vent plug juncture with the cap open.The person in the video wiped the fluid below the vent plug juncture and there was fluid observed on the plum pump as well.The operator stated that when the cap is closed there is no leakage.The reported complaint of leakage on the 14001-31 primary plum set can be confirmed with the cap open.The probable cause is unknown.Without the return of the used sample a comprehensive failure investigation cannot be performed and a cause cannot be determined.The device history review (dhr) for lot 8075764 was reviewed and no non-conformities were found that would have led to the reported complaint.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PRIMARY PLUM SET, 15 MICRON FILTER IN SIGHT CHAMBER, CLAVE SECONDARY PORT, CLAVE
Type of Device
SET, ADMINISTRATION, INTRAVASCULAR
Manufacturer (Section D)
ICU MEDICAL COSTA RICA LTD.
zona franca global
la aurora heredia
CS 
Manufacturer Contact
reed covert
600 n. field dr.
lake forest, IL 60045
2247062300
MDR Report Key18422681
MDR Text Key331672590
Report Number9615050-2023-00826
Device Sequence Number1
Product Code FPA
UDI-Device Identifier10887709100883
UDI-Public(01)10887709100883(17)251001(10)8075764
Combination Product (y/n)N
Reporter Country CodeES
PMA/PMN Number
K141789
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative,Distributor
Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date 12/29/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/29/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number14001-31
Device Lot Number8075764
Was Device Available for Evaluation? No
Date Manufacturer Received11/29/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured10/01/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
UNSPECIFIED FLUID, UNK MFR
-
-